ALTERNATIVE SITE TESTING
IMPORTANT:
There are limitations with AST (Alternative Site Testing).
The AST of the system can be used only during steady-state blood glucose conditions.
Please consult the doctor before you perform AST on the patient.
Measurements from AST should never be used to calibrate a continuous glucose
monitoring systems (CGMs) or entered into insulin dose calculators for insulin dosing
recommendations.
THE PRINCIPAL
SD GlucoNavii™ test strip is designed with an electrode that measures glucose levels.
Glucose in the blood sample mixes with reagent on the test strip which causes a small
electric current. The amount of current that is created depends on how much glucose is
in the blood.
SD GlucoNavii™ meter measures the current that is created and converts the mea-
surement to the amount of glucose that is in the blood. The blood glucose result is
displayed on the meter’s screen. By touching a drop of blood to the tip of the SD
GlucoNavii™ test strip, the strip’s reaction chamber automatically draws the blood into
the strip through capillary action. When the chamber is full, SD GlucoNavii™ meter
starts to measure the blood glucose level. It is a simple and practical system for the
daily monitoring of your blood glucose level.