2. Performing a Control Solution Test
To perform a control solution test you will need the meter, a new test strip and
control solution Level M and/or Level H. The expected control level range is printed on
the test strip vial label.
(Follow your facilities guidelines to determine if M and/or H Lev-
el controls are to be used)
.
NOTE
• A set of Level M and H control solutions is available for purchase. To order
control solutions, contact your supplier.
• Your meter is designed to recognize the difference between the SD Control
solution and blood. Only use SD controls with your GlucoNavii™ meter.
• The meter automatically stores the test results using a control solution code,
letting your review them. Control test results are not included in averages.
STEP-1 :
1) Remove a new test strip from the vial. Be sure to tightly close vial cap after removing
test strip.
2) Insert a test strip (yellow window printed arrow symbol facing up) into test strip slot.
The meter turns on automatically.
STEP-2 :
1) Press the
LEFT
button for 3 seconds to mark the result as a control solution test and
to check the testing system using a control solution in Blood Stand-by Display. If you
don’t want the result marked as a control solution check, press the left button again.
Содержание SD Gluco Navii
Страница 1: ......
Страница 2: ......
Страница 36: ...34...
Страница 37: ...35 Attachment 3 QUALITY CONTROL RESULT RECORD SD GlucoNavii Blood Glucose Monitoring System...
Страница 38: ...36 Attachment 3 QUALITY CONTROL RESULT RECORD SD GlucoNavii Blood Glucose Monitoring System...
Страница 40: ...GlucoNavii Distributed by MAXPERT Medical Expert Products for Expert Care...