27
Blood Sample Error
An insufficient amount of blood was
applied. Discard this test strip and test
again using new test strip and a larger
sample, making sure blood is placed to
the narrow channel in the top edge of
the test strip.
Temperature Error
If the environmental temperature is
above or below the operating range of
a meter, a thermometer icon will
appear on the display. Move to an area
between 50-113°F(10-45°C), wait for
30 minutes, and perform a test. Do not
artificially heat or cool the meter.
Communication Error
The communication between meter
and computer is failed. Reonnect the
meter to the PC.
Internal Error Message
Turn off the meter. Then turn on the
meter again. If there is still an error
message, please contact your supplier.
Strip Error
Defective test strip, used test strip or
the test strip is damaged or inserted
improperly. Discard this test strip and
test again using new test strip.