A. Operator 1 LED:

 Illuminates green when operator 

1 is properly grounded and his/her body voltage level 

is below the preset alarm level. Illuminates solid red 

and audible alarm sounds when operator 1 is above 

the resistance limit. Blinks yellow and audible alarm 

sounds when operator 1 is below the resistance limit. 

Blinks red and audible alarm sounds when the body 

voltage on operator 1 is above the preset alarm level. 

Both the Operator 1 and Operator 2 LEDs blink red 

simultaneously should the operator remote become 

disconnected from the monitor.

B. Mat LED:

 Illuminates green when the worksurface 

mat is properly grounded. Illuminates red and audible 

alarm sounds when the worksurface mat is not properly 

grounded.

C. Operator 2 LED:

 Illuminates green when operator 

2 is properly grounded and his/her body voltage level 

is below the preset alarm level. Illuminates solid red 

and audible alarm sounds when operator 2 is above 

the resistance limit. Blinks yellow and audible alarm 

sounds when operator 2 is below the resistance limit. 

Blinks red and audible alarm sounds when the body 

voltage on operator 2 is above the preset alarm level. 

Both the Operator 1 and Operator 2 LEDs blink red 

simultaneously should the operator remote become 

disconnected from the monitor.

Features and Components

724 Plus Workstation Monitor

D. Power Jack:

 Connect the included 12 VDC power 

adapter here.

E. Set Switch:

 Toggles the monitor's settings for the 

audible alarm, mat monitor circuit, and operator high 

test limit. 

Feature

Procedure

Enable / Disable 

Audible Alarm

1.  Push and release the switch 

while the monitor is powered.

2.  All 3 LEDs will cycle power.

3.  A chirp from the buzzer will 

indicate that the audible alarm 

is enabled. No chirp will indicate 

that the audible alarm is 

disabled.

Enable / Disable 

Mat Monitor 

Circuit

1.  Push and hold the switch until 

all 3 LEDs illmuniate.

2.  Release the switch.

3.  An illuminated MAT LED 

indicates that the mat 

monitor circuit is enabled. A 

non-illuminated MAT LED 

indicates that the mat monitor 

circuit is disabled.

Operator High 

Test Limit -

10 megohms

1.  Disconnect the monitor from 

power.

2.  Push and hold the switch while 

restoring power to the monitor.

3.  The operator LEDs will blink 

green upon boot-up to indicate 

a 10 megohm operator high test 

limit.

Operator High 

Test Limit -

35 megohms

1.  Disconnect the monitor from 

power.

2.  Push and hold the switch while 

restoring power to the monitor.

3.  The operator LEDs will blink red 

upon boot-up to indicate a 35 

megohm operator high test limit.

 

F. Ground Terminal:

 Common ground point for the 

monitor.

G. Monitored Mat Terminal:

 Monitors a worksurface 

mat for proper dissipative resistance. Use the included 

white mat monitor cord to connect the worksurface mat 

to this terminal.

H. Operator Remote Jack:

 Use the included operator 

remote cable to connect an operator remote to this 

connector. The monitor will alarm should the operator 

remote cable disconnect from this connector.

Figure 2. 724 Plus Worstation Monitor features and 

components

A

FRONT

E

BACK

D

B

C

F G

H

Содержание 724 Plus

Страница 1: ...rds shall be established and maintained to provide evidence of conformity to the technical requirements Per ANSI ESD S1 1 Annex A 3 Daily wrist strap system testing may be omitted if constant monitori...

Страница 2: ...wer Jack Connect the included 12 VDC power adapter here E Set Switch Toggles the monitor s settings for the audible alarm mat monitor circuit and operator high test limit Feature Procedure Enable Disa...

Страница 3: ...or will alarm should the operator remote cable disconnect from this connector Workstation 2 Operator Remote Figure 4 Workstation 2 Operator Remote features and components A FRONT BACK C B D A Power LE...

Страница 4: ...1 8 Insert the tinned termination of the ground cord into the ground terminal located on the back of the 724 Plus Workstation Monitor Use a small screwdriver to secure the wire to the terminal block...

Страница 5: ...nitor from power 2 Determine the mounting location of the Workstation 2 Operator Remote Its jack should be accessible to the second operator Locate it within reach of the included operator remote cabl...

Страница 6: ...tor 1 Fit the wristband snugly onto the wrist 2 Snap the wrist cord to the wristband 3 Plug the wrist cord into one of the operator jacks on the Dual Operator Remote The corresponding operator LED wil...

Страница 7: ...240VAC 50 60 Hz Output 12 VDC 1 5 A Output Plug Polarization Center Positive Output Plug 5 5 mm O D x 2 1 mm I D x 9 5 mm L Operator Wrist Strap Dual Wire Amount Monitored 2 Resistance Limit 1 5 mego...

Отзывы: