KeyPad service manual
Rev.1.1
28
Figure 16:
Digi
SET
connector location
Legend:
1.
Digi
SET
connector
5.1.3 Internal wiring connection verification
After removing the bottom plate, verify the connectivity between the
heating block external connector and the PCB.
5.1.4
Digi
SET verification (only if available)
Make sure that the
Digi
SET
is fully functional and that the voltage selector is
properly set (see
Digi
SET
operation and service manual).
5.1.5 SSR and Monostable circuit verification
Fan verification
o
If the SSR case temperature is going too high (more than
85
C) then the fan is not working well. The SSR is probably
also defective. Verify if the fan is running correctly, if not,
then replace the SSR and the fan.