18
CHAPTER 2: INSTALLATION
087-0038 REV L
SCOTT SAFETY
Caution:
When using the sample draw pump, allow sufficient sample time based on
the length of sample tubing you are using. Allow a minimum of 1 second for every foot
of sample tubing and wait for the gas readings to stabilize. Failure to do so could lead
to inaccurate readings.
Installing - Device Powered Down
This section describes how to install the optional pump when the device is powered
down.
1
Connect the pump by inserting, at an angle, the pump's back Tab into the device's
Tab. Rotate the pump forward until the pump sits flush with the device. Tighten the
Thumb Screw into the threaded hole located on the front of the device. See
.
Figure 15
Pump Installation
2
Power on the device and allow it to run through its start-up routine. At this point,
the pump is running.
3
Upon completion of the start-up routine, the pump activates and BLOCK PUMP
displays. To block pump inlet, place a finger over the probe. Once the test is
passed, CLEAR PUMP displays. See
Align Tab
Tighten
Thumb Screw
Содержание Protege
Страница 1: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Страница 2: ......
Страница 3: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Страница 10: ...X LIST OF TABLES 087 0038 REV L SCOTT SAFETY ...
Страница 44: ...20 CHAPTER 2 INSTALLATION 087 0038 REV L SCOTT SAFETY ...
Страница 60: ...36 CHAPTER 4 OPERATION 087 0038 REV L SCOTT SAFETY ...
Страница 76: ...52 CHAPTER 5 MAINTENANCE 087 0038 REV L SCOTT SAFETY ...
Страница 80: ...56 APPENDIX A SPECIFICATIONS 087 0038 REV L SCOTT SAFETY ...
Страница 86: ...62 APPENDIX B QUICK REFERENCE 087 0038 REV L SCOTT SAFETY ...
Страница 90: ...66 APPENDIX C SENSOR INFORMATION 087 0038 REV L SCOTT SAFETY ...
Страница 102: ...78 INDEX 087 0038 REV L SCOTT SAFETY ...
Страница 103: ......
Страница 104: ...Monroe Corporate Center P O Box 569 Monroe NC 28111 Telephone 800 247 7257 FAX 704 291 8330 Web www scottsafety com ...