Maintenance
61
SCOTT SAFETY
087-0048 REV B
Figure 42
IR Connect Software
Hibernate... OK
Clearing a Bump Test
Interval Alarm Alert
This section describes clearing a Bump Test Interval Alarm Alert.
When a monitor is due for a bump test, the monitor toggles between BUP and the
reading. A bump test may be performed at anytime using the Test Station with a target
gas.
Or using the monitor and a target gas, press the front button down once. It displays
numerous screens, then displays the word GAS while the TEST icon flashes. The
monitor waits 45 seconds for the target gas to be applied, or a button press to skip the
bump test. Once gas is applied the check mark appears.
Retrieving Event Logs
and Test Records
Using the USB Pen
Drive
This section describes retrieving the Monitor’s event logs and Test Station test records
using the USB Pen Drive. This allows remote access and verification of the event logs
and test records.
1
Locate the USB Pen Drive metal cover in the upper left hand corner of the Test
Station.
2
Remove the USB Pen Drive metal cover.
3
Remove the USB Pen Drive from the Test Station.
4
Insert the USB Pen Drive into your PC’s USB port. After hardware is found the
Drive Dialog Box appears. See
Figure 43
USB Pen Drive Dialog Box
5
Three (3) folders: Logs, Firmware, and Config are shown, after accessing the USB
via your PC. Only the Log folder applies when accessing the event logs and test
records. Event logs are stored in the ProZM and TS-ProZM folders.
Содержание Protege ZM
Страница 1: ...Protege ZM System Guide PROTEGE ZM PORTABLE GAS SYSTEM 087 0048 Rev B ...
Страница 2: ......
Страница 3: ...Protege ZM System Guide PROTEGE ZM PORTABLE GAS SYSTEM 087 0048 Rev B ...
Страница 10: ...X LIST OF TABLES 087 0048 REV B SCOTT SAFETY ...
Страница 20: ...XX ABOUT THIS GUIDE 087 0048 REV B SCOTT SAFETY ...
Страница 24: ...4 CHAPTER 1 QUICK REFERENCE 087 0048 REV B SCOTT SAFETY ...
Страница 30: ...10 CHAPTER 2 INTRODUCTION 087 0048 REV B SCOTT SAFETY ...
Страница 44: ...24 CHAPTER 3 INSTALLATION 087 0048 REV B SCOTT SAFETY ...
Страница 62: ...42 CHAPTER 4 CONFIGURATION AND SETUP 087 0048 REV B SCOTT SAFETY ...
Страница 98: ...78 CHAPTER 6 MAINTENANCE 087 0048 REV B SCOTT SAFETY ...
Страница 114: ...94 INDEX 087 0048 REV B SCOTT SAFETY ...
Страница 115: ......
Страница 116: ...Monroe Corporate Center P O Box 569 Monroe NC 28111 Telephone 800 247 7257 FAX 704 291 8330 Web www scottsafety com ...