CHAPTER 6: MAINTENANCE
087-0048 REV B
SCOTT SAFETY
Figure 36
Sensor Calibration In Process
5
Remove the monitor from the IR Connect in order to apply gas to the monitor.
6
Apply Calibration Gas to the monitor using Tygon Tubing, Calibration Adapter,
and Regulator prior to the Status Bar timing out.
7
When the monitor’s calibration is successful, all LED’s flash, it beeps, vibrates and
the LCD reflects the reading.
Since the IR Connect is no longer communicating with the monitor, the Status Bar
finishes and the Dialog Box displays Calibration... Time-out.
8
Select CLOSE.