Testing the Monitor
51
SCOTT SAFETY
087-0048 REV B
Figure 34
Bump Test - Calibration Adapter Attached
5
Verify monitor responds to target gas and activates the visual, audible, and
vibrating alarms.
6
Turn off gas cylinder and remove Calibration Adapter.
Warning:
If the monitor fails to activate all alarms within one (1) minute, the monitor
must be taken out of service and tagged. Failure to do so could lead to death or injury.
Bump Testing Using
Test Station
This section describes how to perform a Bump Test using the Test Station.
Equipment required:
This assumes you have already installed the IR Connect Software, and Test Station
Software.
•
Calibration gas – Verify concentration level exceeds set points of monitor and
expiration date of cylinder has not passed.
•
Test Station
•
Monitors
•
IR Connect with USB Cable
•
A PC
1
Ensure an IR WRITE was performed that sent the desired configuration parameters
to the Test Station. See
“Connecting the IR Connect to the Test Station” on
“Configuring the Test Station Software Parameters” on page 36.
2
Ensure the Test Station is powered and the appropriate Gas bottle is installed
properly into the Test Station.
3
Ensure the IR Connect is removed from the Unit 1 slot of the Test Station.
4
Turn power on to each monitor being Bump Tested.
5
Place the desired number of monitors in the four (4) Unit slots.
6
Press the Bump Test Button. Monitors alarm sounds, the Power LED lites Orange
then Green, applicable Unit LEDs lite Orange, while non applicable Unit LEDs lite
Red during the Bump Test. Unit LEDs lite Green when successful or Red if failed.
Содержание Protege ZM
Страница 1: ...Protege ZM System Guide PROTEGE ZM PORTABLE GAS SYSTEM 087 0048 Rev B ...
Страница 2: ......
Страница 3: ...Protege ZM System Guide PROTEGE ZM PORTABLE GAS SYSTEM 087 0048 Rev B ...
Страница 10: ...X LIST OF TABLES 087 0048 REV B SCOTT SAFETY ...
Страница 20: ...XX ABOUT THIS GUIDE 087 0048 REV B SCOTT SAFETY ...
Страница 24: ...4 CHAPTER 1 QUICK REFERENCE 087 0048 REV B SCOTT SAFETY ...
Страница 30: ...10 CHAPTER 2 INTRODUCTION 087 0048 REV B SCOTT SAFETY ...
Страница 44: ...24 CHAPTER 3 INSTALLATION 087 0048 REV B SCOTT SAFETY ...
Страница 62: ...42 CHAPTER 4 CONFIGURATION AND SETUP 087 0048 REV B SCOTT SAFETY ...
Страница 98: ...78 CHAPTER 6 MAINTENANCE 087 0048 REV B SCOTT SAFETY ...
Страница 114: ...94 INDEX 087 0048 REV B SCOTT SAFETY ...
Страница 115: ......
Страница 116: ...Monroe Corporate Center P O Box 569 Monroe NC 28111 Telephone 800 247 7257 FAX 704 291 8330 Web www scottsafety com ...