Select Buffer
1. Press
SET
for 2 seconds to enter setup.
2. Press
▲ to select pH buffer. “BUF” will
appear on the middle of the LCD and
P4.0 will appear on the lower portion.
3. Press
MN/MX/AV
to enter P4.1. The
default
“NIS” (N.I.S.T.) will flash on the
LCD and P4.1 will appear on the lower
portion of the display. If you use N.I.S.T.
buffers, press
MN/MX/AV
to confirm
and the meter returns to P4.0.
4. If your requirement is not for N.I.S.T.
buffers, press
▲ to change the status
to CUSTOM buffer.
5. Press
MN/MX/AV
to confirm and the
meter will return to P4.0.
P5.0 Ready Icon
This feature enables/disables the
“READY” icon, which indicates that
the measured reading is stable.
1. Press
SET
for 2 seconds to enter setup.
2. Press
▲ to select “READY” on
the display.
3. Press
MN/MX/AV
to enter P5.1.
“YES”
will flash on the LCD display and P5.1
will appear on the lower display.
4. Press
▲ to switch between YES or NO.
5. Press
MN/MX/AV
to confirm.
6. Press
ESC
to return to Normal Mode.
P6.0 Temperature Units
To select either Celsius or Fahrenheit temperature scale:
1. Press
SET
for 2 seconds to enter setup.
2. Press
▲ to select “U” on the upper
display. P6.0 is displayed in the lower
portion of display.
8