scigiene RD0370 Скачать руководство пользователя страница 8

 

AVG 

MAX MIN 

Relative Operation

 

When  the  “    

REL

” button is pressed, the  meter will memorize the  present 

reading and the  difference between the  new  reading and the  memorized 
reading and the  result will be  shown on  the  display.  Press the  “     

REL

” 

button again to  exit the  relative  operation.

 

MAX/MIN/AVG Operation

 

 

When  the

 

AVG

 

MAX MIN       

button is pressed, the  meter will enter the

 

MAX/MIN/AVG mode. Under  this  mode the  maximum value,  minimum 

value  and average value  will appear  simultaneously. The  average value  is 

calculated using the  last  8 readings from the  memory, it will update  with 

every  new  reading.

 

 

When  the  MAX symbol is displayed, the  Maximum  value  is shown on  the 

display.

 

 

Press

 

AVG

 

MAX MIN         

again and the  NIN symbol is on  the  display and also  the

 

  minimum  reading.

 

 

Press

 

AVG

 

MAX MIN       

again and the  AVG symbol is on  the  display as  well as  the

 

average reading.

 

 

Press

 

AVG

 

MAX MIN       

again and the  MAX, MIN and AVG will blink together.  This

 

means that  all these readings are  updated in the  memory and the 

reading is the  current temperature.

 

 

Press

 

AVG

 

MAX MIN         

to  circulate the  display mode among these options.

 

 

 

When  the  meter is under 

functions are disabled.

 

AVG 

 

MAX MIN       

operation, the  “     

REL

” and “

°C/°F

 

 

To exit the  MAX/MIN mode, press and hold       

 

 

 

8

 

for two seconds.

Содержание RD0370

Страница 1: ...Digital RTD Thermometer Part RD0370 Instruction Manual...

Страница 2: ...cifications 3 4 Instrument Description 5 6 Operating Instructions 7 9 Temperature Measurement 10 Battery Replacement 11 Calibration Procedure 12 Temperature Meter Calibration Setup 12 14 Product Maint...

Страница 3: ...ith LED backlight C F switchable Min Max Avg functions Relative mode and Data hold Low battery indication with auto power off Includes piercing type temperature probe Pt 100 IEC751 Class A accuracy Id...

Страница 4: ...the following tolerance into the accuracy specification 0 01 of reading 0 03 C 0 01 of reading 0 06 F The basic accuracy Specification does not include the error of the probe refer to the probe accura...

Страница 5: ...Instrument Description 1 Pt type temperature sensor connector 2 LCD display 3 Power ON OFF button 4 HOLD button 5 MAX MIN Average control button 6 Back light button 7 Relative readout button 5...

Страница 6: ...tion Pt xxx Platinum Type Indication HOLD This indicates that the display data is being held MAX The Maximum value is being displayed MIN The Minimum value is being displayed AVG The Average value is...

Страница 7: ...ssing C F button and vice versa back to Celsius Data Hold Operation To hold the present reading and keep it on the display press the HOLD button When the hold data is no longer needed release the data...

Страница 8: ...he last 8 readings from the memory it will update with every new reading When the MAX symbol is displayed the Maximum value is shown on the display Press AVG MAX MIN again and the NIN symbol is on the...

Страница 9: ...disabled Low Battery This meter indicates the battery power level directly on the display Use this reference guide to establish battery power level Battery power level is FULL you can still take meas...

Страница 10: ...t the end of the metal sheath of the sheath type temperature probe To accurately test internal temperatures insert the probe to a distance of at least 15 times the diameter of the sheath directly into...

Страница 11: ...into the instrument 3 Replace cover properly onto the meter and tighten screws The unit s back cover is fitted with rubber rings After replacing the batteries check that the rubber rings are properly...

Страница 12: ...SCCAL 9 Calibrator or https www scigiene com TempCalibrator140 at 0 C in order to calibrate precisely and follow the steps in the calibration setup instructions NOTE regular ice bath methods are not a...

Страница 13: ...en PT probe type Pt100 Pt500 Pt1000 is circulate Press C F button to confirm selection Press and hold Power Light buttons for 2 seconds to enter setup mode Press REL button to enter calibration mode E...

Страница 14: ...2 seconds to enter setup mode Press REL button to enter calibration mode Enter PT selection first Press C F button to confirm PT selection Press Light button to enter recall mode Press C F button to...

Страница 15: ...at contain carbon or benzenes alcohol or anything similar to clean the meter as these substances will damage the surface of the meter Do not use tools with sharp edges screwdrivers metal brushes or an...

Отзывы: