Product Maintenance
In order to ensure the accuracy of the thermometer for a long period of
time it should be calibrated once a year. For service (repairs or calibration)
or additional information on this or any other products, contact Scigiene
at 416-261-4865 or visit our website at: www.scigiene.com.
Clean the device and the window of the display with a clean, lint-free,
antistatic and dry cleaning cloth.
Do no use cleaning agents that contain carbon or
benzenes, alcohol or anything similar to clean the meter as
these substances will damage the surface of the meter.
Do not use tools with sharp edges, screwdrivers, metal brushes or
anything similar to clean the meter.
Temperature Probe
Piercing type
temperature probe
Piercing type temperature probe Specification
Sensor Type
Platinum resistance thermometer sensor Pt 100 (4 wires)
Accuracy
IEC751, class A
±0.15°C ±0.002 measurement temperature
Measurement Range -100 to 400°C
Temp. Sensor Dim.
Approx. 3.2mm ( 0.125”)
Temp. Sensor Length
Approx. 120 mm (4.72”)
Cable Length
Approx. 1100 mm (43.3”)
Water-resistant
EN60529:1991, IP67
15