Operating Instructions

 

 

Turning the meter on

 

 

Press the  

button to  turn  the  thermometer ON or OFF.

 

Connecting the temperature  probe

 

For measurement, plug  the  temperature probe into  the  input  connectors.

 

 

 

Selecting the Temperature  Scale

 

When  the  meter is first turned on,  the  default scale setting is set to  the 

Celsius (°C) scale. It can be  changed to  Fahrenheit (°F) by pressing 

°C/°F

” button  and vice  versa back to  Celsius.

 

Data-Hold  Operation

 

To hold  the  present reading and keep it on  the  display, press the  “

HOLD

”. 

button. When  the  hold  data is no  longer needed, release the  data-hold 

operation by  pressing “

HOLD

” button again.

 

 

When  the  meter is under Data  Hold operation, the  “    

REL

”, 

and “

°C/°F

” buttons are  disabled.

 

Back light Operation 

 

Press the  “

Back Light

” button to  turn  the  back light on, 

press it once again to  turn  it off.  The  meter will turn  the 

back light off automatically if the  “

Back Light

” button is not 

pressed after  10  seconds

 

 

 

7

 

AVG MAX MIN

 

Содержание RD0370

Страница 1: ...Digital RTD Thermometer Part RD0370 Instruction Manual...

Страница 2: ...cifications 3 4 Instrument Description 5 6 Operating Instructions 7 9 Temperature Measurement 10 Battery Replacement 11 Calibration Procedure 12 Temperature Meter Calibration Setup 12 14 Product Maint...

Страница 3: ...ith LED backlight C F switchable Min Max Avg functions Relative mode and Data hold Low battery indication with auto power off Includes piercing type temperature probe Pt 100 IEC751 Class A accuracy Id...

Страница 4: ...the following tolerance into the accuracy specification 0 01 of reading 0 03 C 0 01 of reading 0 06 F The basic accuracy Specification does not include the error of the probe refer to the probe accura...

Страница 5: ...Instrument Description 1 Pt type temperature sensor connector 2 LCD display 3 Power ON OFF button 4 HOLD button 5 MAX MIN Average control button 6 Back light button 7 Relative readout button 5...

Страница 6: ...tion Pt xxx Platinum Type Indication HOLD This indicates that the display data is being held MAX The Maximum value is being displayed MIN The Minimum value is being displayed AVG The Average value is...

Страница 7: ...ssing C F button and vice versa back to Celsius Data Hold Operation To hold the present reading and keep it on the display press the HOLD button When the hold data is no longer needed release the data...

Страница 8: ...he last 8 readings from the memory it will update with every new reading When the MAX symbol is displayed the Maximum value is shown on the display Press AVG MAX MIN again and the NIN symbol is on the...

Страница 9: ...disabled Low Battery This meter indicates the battery power level directly on the display Use this reference guide to establish battery power level Battery power level is FULL you can still take meas...

Страница 10: ...t the end of the metal sheath of the sheath type temperature probe To accurately test internal temperatures insert the probe to a distance of at least 15 times the diameter of the sheath directly into...

Страница 11: ...into the instrument 3 Replace cover properly onto the meter and tighten screws The unit s back cover is fitted with rubber rings After replacing the batteries check that the rubber rings are properly...

Страница 12: ...SCCAL 9 Calibrator or https www scigiene com TempCalibrator140 at 0 C in order to calibrate precisely and follow the steps in the calibration setup instructions NOTE regular ice bath methods are not a...

Страница 13: ...en PT probe type Pt100 Pt500 Pt1000 is circulate Press C F button to confirm selection Press and hold Power Light buttons for 2 seconds to enter setup mode Press REL button to enter calibration mode E...

Страница 14: ...2 seconds to enter setup mode Press REL button to enter calibration mode Enter PT selection first Press C F button to confirm PT selection Press Light button to enter recall mode Press C F button to...

Страница 15: ...at contain carbon or benzenes alcohol or anything similar to clean the meter as these substances will damage the surface of the meter Do not use tools with sharp edges screwdrivers metal brushes or an...

Отзывы: