Battery Replacement

 

 

1)   Remove all of the  back cover screws to  be  able  to  remove the  cover.

 

2)   Verify the  polarity  of the  batteries and install  the  new  LR03 (AAA size)

 

Alkaline  batteries into  the  instrument.

 

3)   Replace cover properly onto the  meter and tighten screws.

 

 

The  unit’s back cover is fitted  with rubber rings.  After 

replacing the  batteries, check that  the  rubber rings  are 

properly placed before reinstalling the  back cover.  Improper 

placement of the  rubber rings  will compromise the  unit’s 

water-resistant structure, and possibly result in damage to 

the  meter.

 

 

 

.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

11

Содержание RD0370

Страница 1: ...Digital RTD Thermometer Part RD0370 Instruction Manual...

Страница 2: ...cifications 3 4 Instrument Description 5 6 Operating Instructions 7 9 Temperature Measurement 10 Battery Replacement 11 Calibration Procedure 12 Temperature Meter Calibration Setup 12 14 Product Maint...

Страница 3: ...ith LED backlight C F switchable Min Max Avg functions Relative mode and Data hold Low battery indication with auto power off Includes piercing type temperature probe Pt 100 IEC751 Class A accuracy Id...

Страница 4: ...the following tolerance into the accuracy specification 0 01 of reading 0 03 C 0 01 of reading 0 06 F The basic accuracy Specification does not include the error of the probe refer to the probe accura...

Страница 5: ...Instrument Description 1 Pt type temperature sensor connector 2 LCD display 3 Power ON OFF button 4 HOLD button 5 MAX MIN Average control button 6 Back light button 7 Relative readout button 5...

Страница 6: ...tion Pt xxx Platinum Type Indication HOLD This indicates that the display data is being held MAX The Maximum value is being displayed MIN The Minimum value is being displayed AVG The Average value is...

Страница 7: ...ssing C F button and vice versa back to Celsius Data Hold Operation To hold the present reading and keep it on the display press the HOLD button When the hold data is no longer needed release the data...

Страница 8: ...he last 8 readings from the memory it will update with every new reading When the MAX symbol is displayed the Maximum value is shown on the display Press AVG MAX MIN again and the NIN symbol is on the...

Страница 9: ...disabled Low Battery This meter indicates the battery power level directly on the display Use this reference guide to establish battery power level Battery power level is FULL you can still take meas...

Страница 10: ...t the end of the metal sheath of the sheath type temperature probe To accurately test internal temperatures insert the probe to a distance of at least 15 times the diameter of the sheath directly into...

Страница 11: ...into the instrument 3 Replace cover properly onto the meter and tighten screws The unit s back cover is fitted with rubber rings After replacing the batteries check that the rubber rings are properly...

Страница 12: ...SCCAL 9 Calibrator or https www scigiene com TempCalibrator140 at 0 C in order to calibrate precisely and follow the steps in the calibration setup instructions NOTE regular ice bath methods are not a...

Страница 13: ...en PT probe type Pt100 Pt500 Pt1000 is circulate Press C F button to confirm selection Press and hold Power Light buttons for 2 seconds to enter setup mode Press REL button to enter calibration mode E...

Страница 14: ...2 seconds to enter setup mode Press REL button to enter calibration mode Enter PT selection first Press C F button to confirm PT selection Press Light button to enter recall mode Press C F button to...

Страница 15: ...at contain carbon or benzenes alcohol or anything similar to clean the meter as these substances will damage the surface of the meter Do not use tools with sharp edges screwdrivers metal brushes or an...

Отзывы: