Viable external triggers:
The following pulses will not trigger the S310:
A laser pulse must be delivered 1 to 3 msec. (depending on the detector type) after an external
trigger occurs or the meter will reset.
Pyroelectric Detector Operation With Oscilloscope
The pyroelectric detectors may be connected directly to a 1 M
Ω
input oscilloscope or a 50 ohm input if the
detectors are the HR type (PHF02, PHF05, PHF09, P05 or P09). The peak voltage of the detector divided by
the V/J or V/mJ output sensitivity = energy (joules or millijoules).
HR Battery Installation
A 9 volt alkaline battery comes installed with all
HR
Series detectors. Always select the OFF position of the
power switch when the detector is not in use to spare the lifetime of the battery. Replace the battery when the
low battery LED indicator lights up.
Remove the two slotted 4-40 binder head screws located on the underside of the detector. Pull off the outer
housing to expose the battery. Remove the used battery from the battery holder and snap in the new battery.
Slide the outer housing back in to place and secure with the screws.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание Vector S310
Страница 4: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 38: ...33 SCHEMATICS Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 39: ...34 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 40: ...35 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 41: ...36 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 42: ...37 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...