7. If The S310 has the optional IEEE488 interface, the REMOTE annunciator will appear along with the
digital interface previously selected; r232 (RS232) or IEEE (IEEE488). Press the SETUP button to toggle
the interface between r232 (RS232) and IEEE (IEEE488). Press the SELECT button when the desired
interface appears in the display.
8. If the optional IEEE488 interface was not installed, the "br" (baud rate) annunciator for setting up the
RS232 interface appears in the display along with the baud rate previously selected. To change the baud
rate press SETUP button until the desired baud rate appears in the display. One of the following baud
rates can be selected: 300, 1200, 2400, 9600, and 19,200. When the preferred baud rate appears in the
display press the SELECT button.
9. The "PA" (parity) annunciator now appears in the display. Press the SETUP button to select none, even,
or odd parity. Press the SELECT button to enter the parity of choice to memory
10. The "HS" (handshake) annunciator now appears in the display. Press the SETUP button until your choice
of none (nOnE), on/off (onoF), or clear to send (CtS) appears in the display. Press the SELECT button
when your choice is displayed to enter the handshake to memory.
11. If IEEE488 has been selected, the "bA" (bus address) annunciator appears with the default bus address.
Bus addresses from 0 to 30 may be selected by pressing the count up (RANGE) count down (MODE)
buttons followed by the SELECT button. Once the remote interface has been setup, the group annunciator
will appear. Press the SELECT button or simply wait a few seconds and the group will automatically be
selected for operation.
Group Configuration for Vector
Pyroelectric Detectors
1. Select the desired group to be configured.
2. Press the SETUP button until the "PYro" annunciator appears. Immediately press the SELECT button to
select the pyroelectric detector configuration program. Either the "V/J" annunciator or the "V/mJ"
annunciator and a detector sensitivity number will appear in the display.
3. Press the SETUP button to toggle between the V/J and V/mJ entries. The output sensitivity of the
pyroelectric detector in V/J or V/mJ is listed on the serial tag of your detector. Use the count up
(RANGE) and count down (MODE) buttons to enter the V/J or V/mJ value. Press the SELECT button to
enter the value to memory. Do not push the SELECT button before entering the sensitivity number or the
next setup stage will be entered.
4. The "SP" (speed) annunciator now appears in the display. Press the SETUP button to toggle between bL
(coated black detector) or hF (high frequency) detector. If the detector Model No. includes the letters
“HF”, press the SELECT button when the hF annunciator appears in the display. If the detector surface
is black (does not have the letters “HF” in the Model No.) press the SELECT button when the "bL"
annunciator appears in the display.
5. The "AUTO, SET CAL" annunciators now appear in the display
(only if watts mode was selected prior to
group configuration)
. This is the transfer calibration function. This allows you to adjust the output
sensitivity of the pyroelectric detector while in operation enabling you to match the average power
reading of the S310 to that of the NIST certified detector. Typically a beam splitter is used with the
detector to be calibrated in one beam path and the NIST certified detector in the other beam path. While
using the S310 with the detector to be calibrated and the NIST standard in an optical calibration, use the
count up (RANGE), count down (MODE) buttons, which change the V/J or V/mJ settings, to make the
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание Vector S310
Страница 4: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 38: ...33 SCHEMATICS Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 39: ...34 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 40: ...35 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 41: ...36 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 42: ...37 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...