11
Range
The RANGE button is used to select the range most appropriate for the power or energy level to be measured.
On the S310 only, held down the RANGE button to toggle the analog meter on or off. The appropriate ranges
will be automatically determined by the type of detector which has been programmed into the group which
has been selected.
1. Press the range button each time to move to the next range. Not pushing the RANGE button when the
desired range appears in the display automatically activates the range after a few seconds and returns the
S310 to its operational state.
2. To turn the analog meter on or off, press and hold the RANGE button, then release after the meter appears
or disappears (S310 only).
Mode
The MODE button allows selection of the type of measurement to be made and units of measure to be
displayed; scientific notation or µ, m annunciators. The different measurement modes available are:
When configured for a pyroelectric detector: Energy (J), Avg. Energy (AVG J), Power (W), and Volts (V).
When configured for a calorimeter: Power (W), Energy (J).
When configured for Ultra detector: Power (W) only.
1. Briefly press the MODE button each time a different mode setting is wanted. Not pushing the MODE
button when the desired mode appears in the display automatically activates that mode after a few
seconds and returns the S310 to its operational state.
2. To select scientific notation or µ and m annunciators, press and hold the MODE button and release when
the desired configuration appears.
Stats
Note:
When using the PHF09 pyroelectric detector in the long pulse setting, a manual range must be
selected when running statistics.
The STATS button provides statistical analysis of a series of energy
readings (pulses). The average energy, minimum energy, maximum energy, standard deviation, and
coefficient of variation are automatically calculated after each run. Runs of up to 1000 pulses (pulse
population) can be entered.
1. To enter into statistics mode press the STATS button. If you are using a pyroelectric detector, the joules
mode and range selected prior to statistics mode are automatically activated. The pulse population
previously selected along with the STATS annunciator and flashing SET annunciator appear in the
display.
2. Press the count up (RANGE) and/or count down (MODE) buttons to change the display to the desired
number of energy readings to be collected. You may select up to 1000 energy measurements for the pulse
population. Press the SELECT button to enter the pulse population to memory. The "STATS"
annunciator will flash indicating the S310 is ready to start a new run.
3. Press the SELECT button to begin collecting new data. The S310 will automatically stop when the pulse
population has been collected.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание Vector S310
Страница 4: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 38: ...33 SCHEMATICS Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 39: ...34 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 40: ...35 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 41: ...36 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 42: ...37 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...