MM-380A
8. Operation Screens
8-17
No.
Item
Description
(2)
Parameter
setting
TRIGGER
/MODE
(Continue
d from
previous
page)
(Continued
to next
page)
MODE: SINGLE TRACE
Force single-trace mode
The instrument measures the maximum
measurement range of current upon input
of a current signal, after which it enters
hold mode. The instrument shows “-” in the
measured value field without showing the
measured value and making any OK/NG
judgment on the value.
MODE: NORMAL TRACE
Force normal trace mode
The instrument measures the maximum
force measurement time upon input of a
force signal. After displaying the data on
screen, the instrument goes into wait state
until the next force signal is input. The
instrument shows “-” in the measured value
field without showing the measured value
and making any OK/NG judgment on the
value.
The maximum current measurement range
varies as follows depending on the TIME and
CURRENT settings in the SETUP (1) screen:
AC
CYC-50/60Hz, CY-***Hz: 4000ms
ms:
2000ms
DC
CYC-50/60Hz: 2000ms
ms:
2000ms
The maximum force measurement range is
6000ms.
*1:
Force when FORCE is selected, and external
±
10V voltage input when External
(EXTERNAL) is selected, for Measurement Selection (SELECT MEAS) in the
STATUS (1) screen.
*2:
Only when FORCE is selected for Measurement Selection (SELECT MEAS) in
the STATUS (1) screen.
*3:
Only when External (EXTERNAL) is selected for Measurement Selection
(SELECT MEAS) in the STATUS (1) screen.