8
schwa-medico
TENS eco 2
Switching ON/OFF the buzzer
Pressing the left key for three seconds, with
E
keyswitching ON/OFF.
Creating a user program (P13 – P24 only)
Choose one program between P13 and P24. Press the
E
key to access the programming
mode. The parameters to be changed flash. You can modify them by pressing the and
keys. Switch to the next parameter by pressing the
E
key. The frequency, the pulse width
and the time can be modified in the user programs.
Factory settings
The parameters from P1 and P13 are identical in new devices, respectively for P2 and P14,
P3 and P15, … P12 and P24.
Switching OFF the TENS unit
Press the key to switch OFF the TENS eco 2. If the battery level is too low or if the TENS
unit has not been used for the previous 2 minutes, it turns off automatically.
Setting date and time (for memory query mode)
Before you can use the memory query mode you need to set the date and time. Pressing
and holding the
E
key for about 3 seconds leads you to the date/time mode. In date/
time mode the LCD shows two figures: The left figure indicates the date, the right figure
indicates the time.
Pressing the left key increases the date by one day (max. 30).
Pressing the left key decreases the date by one day (min. 1).
Pressing the right key increases the time by one hour (max. 23).
Pressing the right key decreases the time by one hour (min. 0).
Press the key to save all parameters and to leave the date/time mode again. Date and
time are continuously updated.
The TENS eco 2 switches OFF automatically after two minutes if no key is pressed. In this
case none of the adjusted parameters are saved.
Memory Query Mode
First set the date and time (see chapter setting date and
time). The memory query mode is used to read out saved
therapy data from patients. By pressing and holding the
right key and
E
key for about 3 seconds you enter the
memory query mode. If no values are saved, the display
shows “null” for about 2 seconds and the device returns
to standby mode automatically. If you have saved therapy
data, the display shows the parameters of the last therapy
session as shown below: