Retrofit kit for fixed scanner
02.00 | SVP-2D / SVP-2D/H / SVP-4 / SVP-3 | Assembly and Operating Manual | en | 0289052
11
■
The RS-232 interface is located on the right side of the housing.
NOTE
The Pin1 assignment of the RS-232 interface differs depending on
the model and software version
DC-DC
converter
DC-DC
converter
Ø
For the SVP-2D / SVP-2D/H with software version 1.3 the
connection cable may not be connected directly to the RS-232
interface.
In this case, a DC-DC converter must additionally be connected
to the RS-232 interface.
Ø
Connect the connection cable, or if required the DC-DC
converter, to the RS-232 interface.
CAUTION
Material damage on the scanner due to improper use of the DC-
DC converter!
•
The DC-DC converter may only be used for TRIBOS SVP3 /
SVP-2D / SVP-2D/H with software version 1.3.
•
If a TRIBOS SVP-2D / SVP-2D/H with software version 1.4 and
SVP-4 is used, connect the connection cable directly to the
RS-232 interface.
NOTE
TRIBOS SVP3 / SVP-2D / SVP-2D/H with software version 1.3
cannot be updated to software version 1.4 or newer versions. The
software version is displayed when the device is siwtched on.