4.4.
Electrical installation / pin assignments
4.4.1.
Power plug
ILME CQM 04/2 or Harting HAN Q 4/2
Pin
Description
Signal
1
Power supply L1
400 V
2
Power supply L2
400 V
3
Power supply L3
400 V
4
Not used
-
PE
Protective conductor
PE
11
Feedback
Potential-free 24V/1A
12
Feedback
Potential-free 24V/1A
CAUTION
The power supply is done via a 7-pin plug, e.g. HARTING HAN Q 4/2 or
ILMA type CQM 04/2.
The plug lock must be locked before start up
(see
figure).
The power supply must comply with the Overvoltage category III or better.
The supply wire must be fused with maximum
16 A
.
Connection PE must be connected on a circuit breaker system which
complies with the safety regulations.
Plug and socket must be in
perfect condition.
1
2
3
PE
11
12
4
Ground connection
POAG
Sensor / actuator plug
Power plug
Plug lock