AR100 Operation Manual
© 2018 Schmitt Industries
Revision 1.1
6
ambient light, lux
Relative humidity
5-95% (no condensation)
Storage
temperature
-
20…+70 °С
Housing material
aluminum
Weight (without
cable)
40 gram
Options to designate when ordering the AR100
Description
1
Red or Blue laser option (450 nm)
2
Measurement range, mm
3
The type of serial interface: 232 (RS232) or 485 (RS485)
4
Analog output of either: 4…20 mA (I) or 0…10 Volts (U)
5
Synchronization input
6
User programmable input/output signal
7
Standard cable length is 2 meter with options for longer cable lengths
Structure and operational principle
Operation of the sensors is based on the principle of optical triangulation
(Figure 1). Radiation of a semiconductor laser (1) is focused by a lens (2) onto an
object (6). Radiation reflected by the object is collected by a lens (3) onto a linear
CMOS array (4). Moving the object (6 - 6') causes the corresponding shift of the
image. A signal processor (5) calculates the distance to the object from the
position of the light spot on the array (4).
Figure 1