Operation Instructions
30 / 42
EN-US · 30.30.01.02533 · 00 · 11/20
6.9 Analog Output Selection
[
F-08
] Select the analog output signal is for flow sensor or pressure sensor.
1. Press
or
button at "Function Selec-
tion Mode" to display [
F-08
] [
ang
].
2. Press
button to switch to "Sensor Selec-
tion".
3. Press
or
button to select the sensor
of analog output (flow sensor or pressure sen-
sor).
NOTE:
1. This function is not available with Output
Specification -02 and -04.
4. Press
button to return to "Function Selec-
tion Mode".
6.10 Accumulated Value Hold Selection
[
F-09
] The default setting is "OFF", the accumulated flow value is zeroed when the power supply is
turned off.
Select this function to keep accumulated flow value to be stored in permanent memory and reload the re-
cent saved accumulated value after power supply turns on.
1. Press
or
button at "Function Selec-
tion Mode" to display [
F-09
] [
eepr
].
2. Press
button to switch to "Accumulated
Value Hold Selection".
3. Press
or
button to turn off the func-
tion or select the data stored cycle.
Function "OFF" = [
off
]
Data stored every 2 min. = [
2Min
]
Data stored every 5 min. = [
5min
]
4. Press
button to return to "Function Selec-
tion Mode".
NOTE:
The maximum writable limit of the memory device is 1 million cycles. If the sensor is operated 24 hours
per day, the durability is calculated as below:
•
5 minutes x 1 million cycles = 5 million minutes = 9.5 years
•
2 minutes x 1 million cycles = 2 million minutes = 3.8 years