![Schiller MAGLIFE LIGHT Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/schiller/maglife-light/maglife-light_user-manual_1212088044.webp)
MAGLIFE light
Part number: 0-48-0080
Page 44
Display
Colour TFT screen: 6.8”; 98 x 132mm; 480 x 640 dots
Connections
SPO
2
and NIBP
Interfaces
RS232, USB 1.1 for connecting the equipment
specified by SCHILLER
Safety standards
IEC
60601-1
EMC
IEC
60601-1-2
CISPER 11 Class B;
with printer connected, class A
The device may be subject to the following
interference without being affected:
Static discharge up to 8 kV
Radio frequency 10 V/m (80 – 2500 MHz, 5 Hz
modulating)
CE marking
According to directive 93/42/EEC class IIb
Protection class
Class I according to IEC 60601-1
5.2 Technical
specifications of modules
5.2.1 Pulse
oximeter
OEM module
BCI
Connection Fibre
optic
Class
CF
SPO2 accuracy
±
2 % from 70 to 99 %
±
3 % from 50 to 69 %
SPO
2
display range
0 – 99%
Pulse accuracy
5 b/min
Pulse display range
30 – 250 b/min
Blocking by defibrillator shock
10 seconds maximum
HF protection
protection from electro surgery devices
5.2.2 NIBP Non-Invasive Blood Pressure
OEM module
CAS
Connection
Fast snap
Class
CF
Measuring principle
Oscillometric
Mode
Manual, Automatic, Continuous
Содержание MAGLIFE LIGHT
Страница 1: ...MAGLIFE light User manual Version 1 2 October 2006 Part No 0 48 0080 ...
Страница 39: ...MAGLIFE light 4 8 Structure of menus 4 8 1 User menus Part number 0 48 0080 Page 39 ...
Страница 40: ...MAGLIFE light Part number 0 48 0080 Page 40 ...
Страница 41: ...MAGLIFE light 4 8 2 Configuration Menus Same for Child Same for Adult Part number 0 48 0080 Page 41 ...
Страница 50: ...11 Front 5 1 3 4 2 6 7 8 9 10 11 12 Part number 0 48 0080 Page 50 ...
Страница 51: ...12 Rear and power supply 13 15 16 17 18 14 19 20 Part number 0 48 0080 Page 51 ...