MAGLIFE light
•
The
Calibration
submenu is used to check if the NIBP pressure module is
measuring the pressure correctly. Connect the external verification system
with an expansion vessel of at least 75 cl and a pressure measuring device.
If required, add an inflating bulb with a deflating valve. Activate
Calibration
;
the display on the screen will show the pressure value measured by the
NIBP module. Make several comparisons between the measurements by
MAGLIFE light
and those by the external measuring device over the entire
measurement range.
Note:
Calibration is regulated in some countries. Comply with
applicable laws and regulations.
Note:
The
Calibration
submenu is locked by a password.
The password is communicated to the people who followed
the technical training.
Part number: 0-48-0080
Page 22