![Schiller CARDIOVIT AT-2 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/schiller/cardiovit-at-2/cardiovit-at-2_service-handbook_1212079024.webp)
Page 1.14
SCHILLER AT-2 6-Channel ECG Unit
SERVICE HANDBOOK Issue 2.1 November 1998
Chapter 1
Operating Elements
Settings (cont.)
Filters
There are three different filters which can be set individually as follows:
Baseline filter
Mains filter
Myogram filter
Baseline Filter
The digital
Baseline filter
suppresses excessive baseline drifts. The setting options are as follows:
Confirm the selection by pressing
STOP
Note:
The set value is the lower limit of the frequency range and is normally set to 0.05
Hz. The settings 0.15 and 0.30 Hz should only be used when absolutely necessary,
as the possibility exists that they could affect the original ECG signal, especially the
ST segments.
Mains Filter
The
Mains filter
is an adaptive digital interference filter designed to suppress AC interference
without attenuating or distorting the ECG.
Set the mains filter in accordance with the frequency of your local mains supply as follows:
Содержание CARDIOVIT AT-2
Страница 10: ...x SCHILLER AT 2 6 Channel ECG Unit SERVICE HANDBOOK Issue 2 1 November 1998 ...
Страница 46: ...Page 3 6 SCHILLER AT 2 6 Channel ECG Unit SERVICE HANDBOOK Issue 2 1 November 1998 Chapter 3 Fault Diagnosis ...
Страница 72: ...Page 5 12 SCHILLER AT 2 6 Channel ECG Unit SERVICE HANDBOOK Issue 2 1 November 1998 Chapter 5 Adjustments ...
Страница 76: ...Page 6 4 SCHILLER AT 2 6 Channel ECG Unit SERVICE HANDBOOK Issue 2 1 November 1998 Chapter 6 Spare Parts ...
Страница 80: ...Page 7 4 SCHILLER AT 2 6 Channel ECG Unit SERVICE HANDBOOK Issue 2 1 November 1998 Chapter 7 Technical Data ...
Страница 88: ...SCHILLER AT 2 6 Channel ECG Unit SERVICE HANDBOOK Issue 2 1 November 1998 Page 8 8 Chapter 8 Glossary ...
Страница 89: ...Page 1 Index SCHILLER AT 2 6 Channel ECG Unit SERVICE HANDBOOK Issue 2 1 November 1998 Index ...
Страница 92: ...Page 4 SCHILLER AT 2 6 Channel ECG Unit SERVICE HANDBOOK Issue 2 1 November 1998 Index ...