8
Unit Settings
8.3
General and unit settings
AT-10 plus
Page 58
A
rt.
-n
o.:
2.5
4004
3 Re
v.
: a
Software
Software Version Display
The current software version for the AT-10 plus. Also displayed is the serial
number of the unit and any options that are installed as follows:
Base configuration (upper case)
M = Measurements (Standard)
C = Interpretation
Options (lower case)
m = memory extension
t = thrombolysis
s = stress
for example Cmt is a C unit (interpretation), with memory extension and
thrombolysis.
Save as default
Save current settings as the default
Restore default
Restore settings to the defined default
Factory default
Restore settings to the factory default
Send default
Store current default setting to a PC
Receive default
Retrieve current default settings from a PC
RS-232
(not
available at time of
print)
Baudrate
This defines the maximum communication rate for the modem (when Modem
is selected in the Mode field (below)).Tick the rate defined in the modem
handbook. Select a Baud rate between 115200 and 14400 Baud, according to
the modem/computer used. Most computers can connect at 115200 Baud and
the standard modem speed is 57600 Baud. If problems are experienced during
transmission reduce the Baud rate.
Mode
Select between
Line
and
Modem
. Line is for when a PC (or other SCHILLER
unit) is connected directly to the RS-232 interface of the AT-10 plus), and
Modem is for transmission over the phone network.
Phone Number
If it is necessary to enter a pause in the dialling sequence enter a comma (,) in
the number. The comma gives a one second pause. If a longer pause is
required, two or more commas can be entered. This may be required for
example, if you have to wait for an outside line. It is also a good idea to insert
a pause after a national or international code.
Modem Init
The code will be found in the user guide for the modem. If the handbook is not
available, the standard Hayes code should work for most modems. This is the
default setting.
The modem initialisation must contain at the minimum, the following
commands with the prefix `AT`.
• `Q0`- modem sends response
• `V0`- numerical response codes
• `E0`- no command echo
• The standard modem initialisation code is: ATB0L1V0Q0E0S0=0
If in doubt about any of these settings, please contact your phone company
and/or modem supplier.
Network
At the time
of print this function
was not fully
defined.
IP Address
These settings must be made or obtained from the network system
administrator.
Parameter
Options
Description