Page 61
Technical Data
9
AT-10 plus
Service Handbook
ECG
9.2
A
rt.
-n
o.:
2.5
4004
3 Re
v.
: a
9.2 ECG
Patient input
Fully floating and isolated, defibrillation-protected (only with original SCHILLER pa-
tient cable)
Leads
• 12 simultaneous leads
• Standard
• Cabrera
Monitor display
Leads
Status
• 3 - 12 channel display of the selected leads
– selectable speed of 5, 10, 20 mm/s
– selectable amplitude 10 or 20 mm/mV
• Filter status (on/off)
• Power source
• Lead selection
• Electrode contact status
• Heart Frequency, HF
• Date and Time
Filters
Myogram filter (muscle tremor)
Line frequency filter
SSF SCHILLER Smoothing fil-
ter
SBS SCHILLER Baseline Stabi-
lizer
• Adjustable at 25 or 35 Hz
• Distortion-free suppression of superimposed 50 or 60 Hz sinusoidal interferences
by means of adaptive digital filtering
Automatic lead programs
3/12-channel presentations of 12 simultaneously recorded leads
Long-term rhythm recordings
1 lead, 10min, 5min or 90sec/page
Exercise ECG with final report
• Automatic control of bicycle ergometer and treadmill (user programmable)
• Final report showing trend plots of heart rate, load and blood pressure, physical
working capacity (PWC 150, PWC 170, PWC max.)
• Options C: QRS and ST measurements
Data record
With optional interpretation (C)
program
• Patient data (name, age, height, weight, BP), user ID
• Listing of all ECG recording conditions (date, time, filter)
• ECG measurements results (intervals, amplitudes, electrical axes)
• Average complexes with optional measurement reference markings
• Guidance on interpreting adult and paediatric ECGs
ECG amplifier
Sampling frequency
Resolution
Pacemaker detection
Frequency range
Measurement range
CMRR
Input Impedance
Defibrillation protection
Patient leakage current
Simultaneous recording of all 9 active electrode signals (= 12 leads)
• 1000 Hz
• 5
µ
V / 12 bit
•
≥
±2 mV /pulse widths
≥
0.1 ms
• 0.05 ... 150 Hz (IEC/AHA)
• dynamic
±
10 mV, DC
±
300 mV
• > 100 dB
• 100 M
Ω
• 5000 VDC
• < 5µA