10. Model preparation

When using hard or hard-elastic material, models are placed into the model cup 

4

, filled up to its upper rim with stainless steel pellets.  When us-

ing soft elastic material, models can be placed on the working platform 

6

.In both cases, the model should be placed with the incisors facing to the 

right towards the locking handle 

3

 or to the left towards the pressure chamber 

2

 to ensure equal thickness of the material between the quadrants.  

11. Scanning/programming by setting code or time

Our pressure moulding material has a 3-digit-code where all required information on temperature, heating time and cooling time is programmed. This 

code is printed in plain text and as bar code on our pressure moulding blanks or foils, except for material with thicknesses below 0.1 and 0.15 mm.  

After selecting your material, press the button 

„code/scan“

.

The scanner 

P

 is now activated. Present only now the bar code of the chosen material to the scanner until you hear a beep. Alternatively, the code 

can be programmed manually.
Program the digits and confirm by pressing the button „code/scan“ again.  You either have the option to enter the heating times manually. Doing so, 

press the key with the clock symbol 

Q

, enter the required heating time and confirm by pressing the button 

Q

 again.

12. Heating

After programming, directly swivel the infrared heater 

1

 over the fixed material. The temperature is regulated automatically by a thermo element.  

The colour of the heater might vary from bright to dark.

13. Pressurising

Before the end of the heating cycle is reached, audible beeping sounds are heard (+ 5 sec. up to 0 sec. = beeping sounds/ 0 sec. up to 5 sec. con-

tinuous beeping sound).The heater should be swivelled back at 0 sec. , at -5 sec. the heater is switched off automatically for safety reasons. Bring 
the pressure chamber 

2

 over the model by turning it 180° to the right

 (make sure the sealing surface is clean)

 and lock it into place by turning 

the locking handle 

C

 180° forwards. Automatically the magnetic valve opens, the pressure chamber fills with air and the cooling time starts, signal 

by blue LED.

14. Depressurising

Optical signals (blue flashing LED) and acoustic signals (4 beeping sounds) indicate the cooling time has elapsed. Press the flashing button 

„air“

 to de-

pressurise the pressure chamber. After approx. 3 seconds 4 beeping sounds indicate the end of depressurising. Now turn the locking handle 

3

 180°  

backwards, push the locking ring 

  to the left and open the pressure chamber 

2

 

. Remove the model. The machine can now be used for 

the next pressure moulding operation. 

15. Special programs using the heating support 

If you want to fabricate positioners, mouth guards or splints, you can manually programme time intervals of 60/120/180/240 seconds. Doing so, the 

pressure chamber has to be closed and the locking handle opened. These intervals are chosen to heat up or to seal the occlusal surfaces using the 

heating support. (REF 3452). After having entered the corresponding heating time, the infra red heater with the heating support hooked in is swivelled 

forwards in the heating position. Depending on the material thickness, this operation can be repeated several times.

16. Service/Maintenance 

The BIOSTAR

®

 machine is maintenance free. However, the air tube including the filter element should be examined regularly. If moisture or debris 

is released, the filter must be replaced. The model cup 

4

 with air holes and model platform 

6

 can be cleaned with an air nozzle or steam jet. The 

reflecting parts of the pressure chamber should be cleaned regularly to obtain the best heating results. For cleaning remove the aluminium adjusting 

ring 

L

 from the pressure chamber 

2

 completely. The sealing surfaces, the sealing ring and the 4 rivets and springs should be cleaned and lightly 

greased also. Place the adjusting ring back on the rivets and springs ensuring that the sealing ring comes to the top and that the whole unit is movable. 

Please also refer to the enclosed „Guideline to solve technical problems“.

17. Service updates

You‘ll receive a registration card with the machine. By signing-up, you‘ll be provided automatically with information on updates. You can download the 

updates on our website 

www.scheu-dental.com/downloads/support

Extract the file to an empty and formatted removable USB device. Connect the USB device to the the 

service interface

 

S

 on the back of the machine. 

Start the machine, then press 5 times the 

button „OK“

. Now you‘ll be asked to enter the pass word for the service menu.

Enter the 

figure „1“

 and confirm by pressing the

 button „OK

“. By pressing agin the 

figure „1”

 the update starts. After the update is finished, quit 

the service menu by pressing the 

button „CE+home“

.

Technical modification, which serve for improvement, are subject to our agreement. Reproduction, even in extracts, is only allowed with our written approval.

Содержание BIOSTAR

Страница 1: ...Bedienungsanleitung Operation Manual Mode d emploi Manual de Instrucciones Istruzioni per l uso ...

Страница 2: ...t dienen vorbehalten Nachdruck auch auszugsweise nur mit unserer schriftlichen Genehmigung Technical modification which serve for improvement are subject to our agreement Reproduction even in extracts is only allowed with our written approval ...

Страница 3: ...nical data Spannung AC 230 V 115 V 100 V 50 60 Hz Leistung Power 750 VA B x H x T W x H x D 450 x 230 x 260 mm Gewicht Weight 14 kg Druck Pressure max 6 bar 87 psi J J 9 R 9 7 R 8 7 230 V 115 V 100 V Abb Fig 1 Abb Fig 2 Abb Fig 3 Abb Fig 4 S T 21 22 N ...

Страница 4: ...den Umwelt Das Gerät ist aus wiederverwendbaren oder recyclebaren Materialien hergestellt Gemäß der 2012 19 EU Elektro und Elektronik Altgeräte Richtlinie nimmt SCHEU DENTAL das Gerät bei freier Rücklieferung kostenlos zurück Alternativ kann es an einer Sammelstelle für das Recycling elektrischer und elektronischer Geräte abgegeben werden 2 Allgemeine Inbetriebnahme Zur optimalen Nutzung des BIOST...

Страница 5: ...ckdose 7 auf der Geräterückseite verbinden Bei der Geräteausführung in 230 V befinden sich in der Netzsteckdose auch zwei Gerätesicherungen die die Elektronik vor Überspannungen schützen Zum evtl Austausch drücken Sie bitte den kleinen Clip unter der Netzsteckdose und entnehmen Sie das Sicherungsfach 8 Bei der Geräteausführung in 100 V 115 V befinden sich die Sicherungen in 2 separaten Halterungen...

Страница 6: ...Jetzt Verschlusswelle 3 um 180 nach hinten drehen Verschlussring K nach links schieben und Druckkammer 2 öffnen Nach Entnahme des Modells ist das Gerät für den nächsten Tiefziehvorgang wieder bereit 15 Sonderprogramme für den Heizsupport Bei geschlossener Druckkammer und geöffneter Verschlusswelle können speziell bei der Herstellung von Positionern Mundschutz oder Schienen Zeitintervalle von 60 12...

Страница 7: ...ith the directive 2012 19 Waste Electrical and Electronic Equipment legislation WEEE SCHEU DENTAL accepts to take back the device under the condition of free return delivery Alternatively the machine can be sent to a designated collection point for the recycling of waste electrical and electronic equipment 2 General commissioning Please make sure to understand and follow the below instructions to ...

Страница 8: ...e to the corresponding network by plugging in the original power cable on the back of the machine 7 please also refer to connection to the local network in the safety instructions Two fuses are built in to protect the machine from high voltage 230 V version To exchange the fuses press the little clip below the connector 8 and take off the fuse For the 100 V 115 V versions these fuses are built in ...

Страница 9: ... seconds 4 beeping sounds indicate the end of depressurising Now turn the locking handle 3 180 backwards push the locking ring K to the left and open the pressure chamber 2 Remove the model The machine can now be used for the next pressure moulding operation 15 Special programs using the heating support If you want to fabricate positioners mouth guards or splints you can manually programme time in...

Страница 10: ...de protection Environnement L appareil est fabriqué à partir des matériaux réutilisables ou recyclables Conformément à la directive relative aux déchets d équipements électriques et électroniques DEEE 2012 19 SCHEU DENTAL accepte de reprendre l appareil sous réserve que le retour est payé par l expéditeur Alternative ment l appareil peut être envoyé à une décharge spécialisée pour la récupération ...

Страница 11: ...aide du câble fourni à la prise au courant 7 au dos de l appareil Pour les modèles en 220 V il y a deux fusibles miniatures dans la prise au courant au dos de l appareil protégeant l électronique contre la surtension Pour les changer si nécessaire presser la petite agrafe en dessous de la prise au courant et enlever les fusibles 8 Pour les modèles en 100 115 V ces deux fusibles se trouvent dans de...

Страница 12: ...ès environ 3 secondes annoncent la fin du procédé de décompression Tournez l arbre de serrage 3 de 180 en arrière pousser le cercle de serrage K à gauche et ouvrir la chambre de pression 2 Après avoir enlevé le modèle l appareil est prêt à démarrer un nouveau procédé de thermoformage 15 Programmes spécifiques en utilisant le support de chauffe Si vous voulez fabriquer des positionneurs des protège...

Страница 13: ...e El aparato se ha fabricado con materiales reutilizables o reciclables Conforme a la directiva sobre residuos de aparatos eléctricos y electrónicos 2012 19 UE SCHEU DENTAL acepta la devolución voluntaria del aparato de forma gratuita Cabe también la posibilidad de entregar los aparatos eléctricos y electrónicos en un punto de recogida para su reciclaje 2 Puesta en funcionamiento general Es precis...

Страница 14: ...la parte posterior del aparato La versión de 230 V del aparato dispone de dos fusibles en la toma de alimentación que protegen el sistema electrónico frente a subidas de tensión En caso de que sea necesario cambiarlos pulse el pequeño cierre situado bajo la toma de alimentación y saque el compartimento del fusible 8 En las versiones de 100 V 115 V los fusibles se encuentran en 2 sujeciones separad...

Страница 15: ...el eje de bloqueo 3 180 hacia atrás desplace el anillo de cierre K hacia la izquierda y abra la cámara de presión 2 Tras la extracción del modelo el aparato está listo de nuevo para el siguiente proceso de embutición profunda 15 Programas especiales para apoyo del calentamiento Con la cámara de presión cerrada y el eje de bloqueo abierto se pueden programar intervalos de 60 120 180 240 segundos pa...

Страница 16: ...ettiva 2012 19 UE sui rifiuti di apparecchiature elettriche ed elettro niche SCHEU DENTAL ritira senza alcun costo l apparecchio usato con modalità di restituzione gratuita In alternativa esso può essere conferito a un centro di raccolta per il riciclaggio di apparecchiature elettriche ed elettroniche 2 Istruzioni generali per la messa in esercizio Per l utilizzo ottimale di BIOSTAR è necessario l...

Страница 17: ...esa di rete 7 sul retro dell apparecchio Nella versione dell apparecchio a 230 V nella presa di rete si trovano anche due fusibili che proteggono l elettronica dell apparecchio dalle sovratensioni Per l eventuale sostituzione premere la piccola clip sotto la presa di rete ed estrarre lo scomparto portafusibili 8 Nella versione dell apparecchio a 100 V 115 V i fusibili sono collocati in 2 supporti ...

Страница 18: ...po circa 3 secondi indicano la fine del processo di sfiato Ora ruotare l albero di chiusura 3 di 180 verso il retro spingere l anello di chiusura K verso sinistra e aprire la camera di pressione 2 Dopo la rimozione del modello l apparecchio è nuovamente pronto per il successivo processo di termoformatura 15 Programma speciale per il supporto per il riscaldamento Con camera di pressione chiusa e al...

Страница 19: ...ingefettet ist Nur so werden ein leichtes Verriegeln und ein geringer Verschleiß der Dichtung gewährleistet Eine verschlissene Dichtung kann durch einfaches Herausziehen aus der Hülsen halterung beim BIOSTAR getauscht werden Vor dem Einsetzen einer neuen Dichtung sollten Sie auch die äußere Seitenfläche der Dichtung einfetten Bild 4 In Ihrem Zubehör Paket finden Sie unter anderem 3 Ersatzdichtunge...

Страница 20: ...schmutzung nicht mehr gängig sein Pflegehinweis Bei der Durchführung von nebenstehenden Pflegemaßnahmen am Tiefziehgerät achten Sie bitte darauf dass die Beweglichkeit des Ausgleichsrings jederzeit sichergestellt ist Der Ausgleichsring ist durch Eindrücken an einer Seite und gleichzeitiges Anheben auf der gegen überliegenden Seite aus der Druckkammer herauszu nehmen Auch die 4 Nieten mit den Druck...

Страница 21: ... well due to debris Maintenance hint If you follow the maintenance advices given on the opposite side you ensure the mobility of the balancing ring Following the main tenance advices given on the opposite side please make sure the balancing ring is mobile at any time Pull out balancing ring of pressure chamber by pressing on one side and lifting up on the opposite side at one time Now the 4 nuts w...

Страница 22: ...uve à modèle L anneau d ajustement peut être penché ou bien bloqué dû à des salissures Astuce de maintenance Veiller à ce que la mobilité de l anneau d ajustement soit assurée à tout moment en effectuant les mesures de maintenance mention nées ci contre Retirer l anneau d ajuste ment de la chambre de pression en pressant un côté et en même temps soulevant l autre côté Retirer maintenant les 4 rive...

Страница 23: ... de suciedad y está bloqueado Instrucción de manutención Fijarse en la movilidad del anillo de compensación en todo momento durante la ejecución de las medidas de manutención contiguas Sacar el anillo de compensación de la cámara de presión calcando un lado y levantando el de enfrente Ahora también se pueden sacar los 4 remaches con los muelles de compresión Limpiar y engrasar con vaselina las sup...

Страница 24: ...compensatore è inclinato o bloccato causato da insudicia mento Consiglio per la manutenzione Effettuando di tanto in tanto manutenuzione descritta qui a fianco l anello compensatore manterrà sempre la sua mobilità Premendo l anello compen satore da un lato esso si solleva all altro ed è facilmente estraibile dalla muffola Anche i quattro ribattini con le molle a compressione devono essere estratti...

Страница 25: ... une manipulation inappropriés Les cas de non respect des consignes de manipulation nettoyage et maintien Les cas de branchement inapproprié ou d installation incorrecte Les cas de force majeure ou d autres influences externes Les cas de modifications effectuées sur l appareil Les cas d utilisation des pièces détachées d un autre fabriquant Les cas de signes d usures dus à l âge n ayant pas d effe...

Страница 26: ...ly tested acc to the International Guidelines ISO 10993 and EN ISO 7405 with the result that it has been released for clinical use F Nos plaques sont été contrôlées sur le plan de la biologie cellulaire selon les Normes Internationales ISO 10993 et EN ISO 7405 étant approuvées pour l utilisation clinique E Nuestras planchas según las normas internacionales DIN EN ISO 10993 y EN ISO 7405 y son perm...

Отзывы: