Guideline to solve technical problems
Error
Probable reason
Solution
The foils are partially burned or too soft
You chose a wrong code or hea
ting-time for the foil.
Please consider the hints and da
ta given on the label of each
packa
ge.
Pressure moulding results are not sa
tisfying
You chose a wrong code or hea
ting-time for the foil.
Please consider the hints and da
ta given on the label of each
packa
ge.
No pressure build-up
When embedding the model in pellets,
the model cup has not been
filled up to the c
losing area.
So the foil slips a
way from the foil-fixa
tion because it is extended
too much.
Under these circumstances a correct sealing is not given.
Please make sure tha
t only foils with a diameter of 125 mm are
used!
Fill the model cup with
pellets up to the c
losing
area.
Balancing ring is not straight or not running well due to debris.
Maintenance hint:
If you follo
w the maintenance advices given on the opposite side
you ensure the mobility of the balancing ring.
F
ollo
wing the main
-
tenance advices given on the opposite side,
please make sure the
balancing ring is mobile a
t an
y time.
Pull out balancing ring of
pressure chamber by
pressing on one side and
lifting up on the opposite
side a
t one time.
No
w the 4 nuts with
pressure springs can be
pulled out.
After c
leaning
and greasing with vaseline,
reinsert ring.
Difficulties in c
losing pressure chamber
Pellets on the c
losing area of housing.
Before c
losing the pressure chamber make sure tha
t the c
losing
area is c
lean!
Difficulties in opening pressure chamber
The different steps for opening the pressure chamber ha
ve not been
carried out correctly
.
After pushing the air button
for depressuriza
tion and
waiting for 5 seconds the
follo
wing opera
tions ha
ve
to be done:
1
Turn the locking handle
180° to the back
2
Turn the bayonett c
losing ring to the left
3
Open the pressure chamber
1
2
3