Portable analyzer
9 of 18
GA-S1
Version 2015
6. Calibration
Please carry out the calibration only outdoors, since the effluent gas can be a health risk for
every person located in a closed room.
Like in chapter 5: Control procedure you have to use calibrating gas to calibrate the analyzer. During
calibration two calibration points are set for both sensors (O
2
and CO
2
), “low” and “high”.
Sensor
Point
Used medium
Value
O
2
Low
Calibrating gas (i.e. nitrogen)
0,0%
High
Fresh air
20,9%
CO
2
Low
Fresh air
0,04%
High
Calibrating gas (CO
2
)
5,0%
From these calibration points a linear gradient emerges which offers a reliable measurement.
For a reliable measurement, please observe the minimum warm-up time in the lower right
corner of the display until
“READY” is displayed in the lower left corner.
6.1 Calibrating the O
2
sensor with zero % - gas (nitrogen)
Open calibration menu by pressing function key F2 (CALIB)
For O
2
sensor press function key F2 (O2)
Press function key F3 (REF 0%)
Connect nitrogen (zero gas O
2
=0% N
2
=100%), see image 1 or 2
Press function key F4 (NEXT)
o
Calibration can be interrupted by pressing function key F1 (EXIT)
If the measured value is stable, save the value with function key F4 (SAVE)
To stop the calibration press function key F1 (EXIT)