1.  Introduction 

This user manual applies to the following company group:  

 

Schele Electronic UG & Co KG 

1.1  About this manual 

This manual contains instructions about how to work with the portable analyzer of Schele Electronic in 
order to lead to safe and efficient operations. 
Please read the entire manual carefully before starting up the portable analyzer. 

1.2  Explanation of the symbols 

In this user manual, this symbol stands for all instructions and regulations related to 
your safety and the safety of any third party. 

 

 

You find this symbol at important instructions and information for trouble-free operations and 
a correct usage of the portable analyzer. 

 

1.3  Ranges of measurement 

0-25% O

2

 and 0-10% CO

2

 

 
or 
 
0-25% O

2

 and 0-30% CO

2

 

 
The range of measurement is dependent on the type of analyzer you purchased. 
 

The portable analyzer is meant to measure oxygen (O

2

) and/or carbon dioxide (CO

2

in cold stores for fruits and vegetables. The O

sensor type used is not suitable for 

cold stores where the CO

2

 is removed through chalk. 

This O

2

 sensor is particularly suitable for stores treated with an active carbon 

scrubber.

 

 
Attention

Hydrocarbons can influence the measurement of the oxygen sensor. 
These hydrocarbons are not removed by chalk. 

Содержание GA-S1

Страница 1: ...User manual Copyright 2015 Schele Electronic UG Co KG Portable analyzer GA S1 for O2 and CO2 Version 2015 5...

Страница 2: ...NGES OF MEASUREMENT 3 1 4 CONTACT 4 2 GUARANTEE 4 3 HANDLING 4 4 ANALYZER START UP 5 4 1 CHECKING THE ANALYZER 6 5 CONTROL PROCEDURE 7 6 CALIBRATION 9 6 1 CALIBRATING THE O2 SENSOR WITH ZERO GAS NITRO...

Страница 3: ...e safety of any third party You find this symbol at important instructions and information for trouble free operations and a correct usage of the portable analyzer 1 3 Ranges of measurement 0 25 O2 an...

Страница 4: ...he portable analyzer appropriately as described in this user manual the guarantee period is 12 months 3 Handling ON Switch on Push 2 seconds OFF Switch off Function keys F0 Changing company name at st...

Страница 5: ...is included in the delivery A new hydrophobic filter can be ordered separately see Table 2 Spare parts After warming up the screen will display the current O2 and or CO2 measurement CO2 trend upwards...

Страница 6: ...ng gas see chapter 5 Control procedure If the given values on the display do not match with the values of the calibrating gas calibration is necessary see chapter 6 Calibration In the event of no samp...

Страница 7: ...the analyzer can suck in This can be identified by gas escaping the open connector Alternatively a small plastic or glass bottle can be used from which the analyzer can draw a gas sample through anoth...

Страница 8: ...MAX 2L min Aansluiten op handmeter Connection of portable analyser Anschlu am tragbare Me ger te Raccordement S url analyseur portable Image 2 Connection of calibrating gas with a testing bottle Conne...

Страница 9: ...as CO2 5 0 From these calibration points a linear gradient emerges which offers a reliable measurement For a reliable measurement please observe the minimum warm up time in the lower right corner of t...

Страница 10: ...1 or 2 Press function key F4 NEXT o Calibration can be interrupted by pressing function key F1 EXIT If the measured value is stable save the value by pressing function key F4 SAVE To stop the calibra...

Страница 11: ...Portable analyzer 11 of 18 GA S1 Version 2015 If the measured value is stable save the value with function key F4 SAVE To stop the calibration press function key F1 EXIT...

Страница 12: ...rrent measurement is displayed Press the right arrow key twice until the following screen is displayed Press function key F2 SAVEVAR Confirm default save with function key F4 YES or cancel the process...

Страница 13: ...king correctly you are able to reset the analyzer to factory settings Important After reset the analyzer has to be recalibrated for reliable measurements Make sure the screen with the current measurem...

Страница 14: ...ry is always fully charged During a long time of inactivity charge battery onto 40 and recharge it to this level every 2 to 3 months This avoids a deep discharge of the battery which could lead to a b...

Страница 15: ...ect polarity of the battery and avoid short circuits especially involving non insulated connector systems Do not allow the battery to contact water or any other fluid Do not charge the battery with re...

Страница 16: ...detectable loss of performance To avoid further discharging disconnect the pack from the consumer unit after use Memory effect cell capacity Since LiPo cells do not suffer from the memory lazy batter...

Страница 17: ...ime Make sure the screen with the current measurements is displayed Press function key F4 SCREEN The operating time is displayed in h mm ss the number of days and total number of hours Return to curre...

Страница 18: ...ng temp 20 70 C LCD Display 70 x 38 pixels Air flow max 0 4 l min Nylon filter 0 45 m 30mm PVC tube D 3mm 5mm Pump Electronics control Case aluminum L 260 x B 150 x H 60 mm Battery Nexspor NXP 1120 5C...

Отзывы: