Portable analyzer 

           5 of 18 

 

GA-S1

 

 

Version 2015

 

4.  Analyzer start up 

The portable analyzer is switched on at 

the “ON” button (Push for 2 seconds). 

At the same time an air pump is switched on. This pump is responsible for the suction of 
the gas sample. 
 

For a reliable measurement, please observe the minimum warm-up time in the 
lower right corner of the display until the word 

“READY” is displayed in the lower left corner 

  

 
 
Since the analyzer sucks in the gas sample autonomously, do not insert gas under pressure into the 
analyzer. The gas sample must not contain any water at all. 
To protect the sensors a dust/dirt/moisture filter is included in the delivery. 
A new hydrophobic filter can be ordered separately, see Table 2: Spare parts. 

 

 

 
 
 

 
 
 
 
 
After warming up the screen will display the current O

2

 and/or CO

2

 measurement: 

 
CO2 

 = trend upwards 

O2 

 = trend downwards 

 
READY = ready for use 
OK 

Battery full indication 

LA 

Battery low indication 

 
0: 4: 23  =  Operating time [h:mm:ss] is 0 hours 4 
minutes 23 seconds 
 

Never work with the analyzer without this hydrophobic filter. 

 

 

The analyzer has to be recharged with the battery charger included in the 
delivery. Take care that the built-in battery is always completely recharged. 
Charge battery only at a temperature higher than 5 °C. 
During a long time of inactivity, charge battery onto 40% and recharge it to 
this level every 2 to 3 months. This avoids a deep discharge of the battery 
which could lead to a breakdown of the battery. 

Содержание GA-S1

Страница 1: ...User manual Copyright 2015 Schele Electronic UG Co KG Portable analyzer GA S1 for O2 and CO2 Version 2015 5...

Страница 2: ...NGES OF MEASUREMENT 3 1 4 CONTACT 4 2 GUARANTEE 4 3 HANDLING 4 4 ANALYZER START UP 5 4 1 CHECKING THE ANALYZER 6 5 CONTROL PROCEDURE 7 6 CALIBRATION 9 6 1 CALIBRATING THE O2 SENSOR WITH ZERO GAS NITRO...

Страница 3: ...e safety of any third party You find this symbol at important instructions and information for trouble free operations and a correct usage of the portable analyzer 1 3 Ranges of measurement 0 25 O2 an...

Страница 4: ...he portable analyzer appropriately as described in this user manual the guarantee period is 12 months 3 Handling ON Switch on Push 2 seconds OFF Switch off Function keys F0 Changing company name at st...

Страница 5: ...is included in the delivery A new hydrophobic filter can be ordered separately see Table 2 Spare parts After warming up the screen will display the current O2 and or CO2 measurement CO2 trend upwards...

Страница 6: ...ng gas see chapter 5 Control procedure If the given values on the display do not match with the values of the calibrating gas calibration is necessary see chapter 6 Calibration In the event of no samp...

Страница 7: ...the analyzer can suck in This can be identified by gas escaping the open connector Alternatively a small plastic or glass bottle can be used from which the analyzer can draw a gas sample through anoth...

Страница 8: ...MAX 2L min Aansluiten op handmeter Connection of portable analyser Anschlu am tragbare Me ger te Raccordement S url analyseur portable Image 2 Connection of calibrating gas with a testing bottle Conne...

Страница 9: ...as CO2 5 0 From these calibration points a linear gradient emerges which offers a reliable measurement For a reliable measurement please observe the minimum warm up time in the lower right corner of t...

Страница 10: ...1 or 2 Press function key F4 NEXT o Calibration can be interrupted by pressing function key F1 EXIT If the measured value is stable save the value by pressing function key F4 SAVE To stop the calibra...

Страница 11: ...Portable analyzer 11 of 18 GA S1 Version 2015 If the measured value is stable save the value with function key F4 SAVE To stop the calibration press function key F1 EXIT...

Страница 12: ...rrent measurement is displayed Press the right arrow key twice until the following screen is displayed Press function key F2 SAVEVAR Confirm default save with function key F4 YES or cancel the process...

Страница 13: ...king correctly you are able to reset the analyzer to factory settings Important After reset the analyzer has to be recalibrated for reliable measurements Make sure the screen with the current measurem...

Страница 14: ...ry is always fully charged During a long time of inactivity charge battery onto 40 and recharge it to this level every 2 to 3 months This avoids a deep discharge of the battery which could lead to a b...

Страница 15: ...ect polarity of the battery and avoid short circuits especially involving non insulated connector systems Do not allow the battery to contact water or any other fluid Do not charge the battery with re...

Страница 16: ...detectable loss of performance To avoid further discharging disconnect the pack from the consumer unit after use Memory effect cell capacity Since LiPo cells do not suffer from the memory lazy batter...

Страница 17: ...ime Make sure the screen with the current measurements is displayed Press function key F4 SCREEN The operating time is displayed in h mm ss the number of days and total number of hours Return to curre...

Страница 18: ...ng temp 20 70 C LCD Display 70 x 38 pixels Air flow max 0 4 l min Nylon filter 0 45 m 30mm PVC tube D 3mm 5mm Pump Electronics control Case aluminum L 260 x B 150 x H 60 mm Battery Nexspor NXP 1120 5C...

Отзывы: