Portable analyzer
10 of 18
GA-S1
Version 2015
6.2 Calibrating the CO
2
sensor with reference gas
Open calibration menu by pressing function key F2 (CALIB)
For CO
2
sensor press the function key F3 (CO2)
Press function key F3 (REF %)
o
If necessary, adjust the CO
2
reference value with function key F3 (SETVAR) and the arrow
keys
o
When the value is changed, close with key RET
Connect the reference gas, see Image 1 or 2
Press function key F4 (NEXT)
o
Calibration can be interrupted by pressing function key F1 (EXIT)
If the measured value is stable, save the value by pressing function key F4 (SAVE)
To stop the calibration press function key F1 (EXIT)
6.3 Calibration of fresh air
Open calibration menu by pressing function key F2 (CALIB)
Press function key F4 (F.AIR)
If you use the included hose:
o
Close the valve of the calibrating gas
If you use the testing bottle:
o
Take out the hose connected to the analyzer of the bottle
Press function key F4 (NEXT)
o
Calibration can be interrupted by pressing function key F1 (EXIT)