18
Proscan 2200 (Chromatic Sensor) User Manual V.4.0
8.0
Routine Maintenance
Scantron recommend an annual system Verification and routine maintenance regime be
adopted.
Part
Description
Period
Z axis lead-screw
Apply THK multipurpose lithium EP grease
to lead-screw
Every 12 months
X and Y rails
Apply ISO68 Slide-way oil to rails
Every 12 months
Sensor measuring
head optics
Make sure laser optics are clean and free
from dust. Clean using a lint free wipe and a
small amount of methanol
Every 3 months
\after handling \if
data quality falls
X and Y axis linear
encoder strip and
readhead
Use a clean dry lint free cloth.
Every 12 months \if
readhead shows red
Use only the following solvents sparingly
with a wetted lint free cloth
-
Propan 2 ol (iso propyl alcohol)
CH
3
CHOHCH
3
-
N Heptane CH
3
(CH
2
)
5
CH
3
DO NOT
use the following aggressive
solvents to clean the scale
-
Acetone CH
3
COCH
3
-
Chlorinated solvents
-
Benzene
-
Methylated spirits
Cables
Check for wear\breakages
Every 12 months\as
necessary