SCALES usa © 2005
Page
14 of 16
OPTION SETTINGS
To set the optional scale technical parameters, it is necessary to enter
a setup menu. This is done by entering a key combination while the
scale is in the weighing mode.
To enter the settings menus press and hold the
CE
key and then
briefly press the
QUANTITY CHECK
key and release both keys. The
parameter menu has 7 functions that can be toggled thru using the
ZERO
key The
Weight Display
will show the name of the functions
and the
Unit Weight Display
will show the setting.
To step through available settings within a parameter, press the
ZERO
key.
At any time. Press the
UNIT WEIGHT / CANCEL
key to return to
weighing.
SETUP TABLE 1
Weight
Display
Description and Settings (Shown in Unit Weight Display)
FILt ?
). SELECTION OF FILTER GRADE.
There are two filter modes
(mode 1 and mode 2) for the scale. Each mode has four grades (0-3). The
higher the grade is, the higher the stability is but the slower the reaction
is. For example,
“
SEL 1 0
”
means the reaction of the scale is the
quickest, but its stability is the lowest.
“
SEL2 3
”
means the reaction of
the scale is the lowest, but its stability is the highest.
In the
FILt ?
menu, press
M+
to enter selection of filter grade mode.
Then step thru the grade modes using the
ZERO/>
key until you reach the
desired mode. Press
M+
to save it. The display will show next menu. Or
press the
UNIT WEIGHT/CANCEL
key to exit.
StAb ?
SELECTION OF STABILITY.
There are five stable grades(0
—
4). In
the
“
StAb ?
”
menu, press
M+
to enter it. Then choose the grade by shift
key
ZERO/>
. The original grade is 1. The higher the grade is, the higher
the stability is. Press
M+
to save it. The display will show next menu. Or
press the
UNIT WEIGHT/CANCEL
key to exit.
2
trC ?
SELECTION OF ZERO TRACING GRADE.
There are five zero
tracing grade (0-4). In the
“
2
trC
?
”
menu, press
M+
to enter it. Then
choose the grade by shift key
ZERO/>
. The original grade is 1. The
higher the grade is, the larger the zero tracing is. . Press
M+
to save it.
The display will show next menu. Or press the
UNIT
WEIGHT/CANCEL
key to exit.