4
b)
Tare cancellation:
to eliminate the stored tare, press the
[
Tare/Zero
]
key. The tare weight is cancelled and the gross
weight is displayed. The tare light turns off.
III. CALIBRATION
(1)
Enter in the calibration menu:
press the
[Clear/CAL]
key for about 3 seconds: the instrument shows the A/D
converter points relative to the weight.
(2)
press the
[Clear/CAL]
key
,
the display
shows
“F= xxx”
: select the maximum capacity, with
[
Set/Div
]
key;
(3)
press the
[Clear/CAL]
key
,
the display
shows
“d=0.xx”
:
select the divisions of the scale, with
[
Set/Div
]
key;
(4)
press the
[Clear/CAL]
key
,
the display
shows
“CAL-0”
(zero calibration): clear the loading and wait for stable (the
arrow on “Stable” turned on).
(5)
press the
[Clear/CAL]
key
,
the display
shows
“L-xx”
(calibration with sample weight): press the
[
Set/Div
]
key for select
the weight calibration;
(6)
Load the scale and wait the stable.
(7)
Press the
[Clear/CAL]
key for save and exit.
Note: during the calibration, press the [Tare/Zero] key to return to the previous step or exit the calibration; if one
exits before the last step, the modifications will not be saved; only when one ends with step (7), the parameters are
updated and the modifications are saved.
IV. BACKLIGHT FUNCTION
1) Press the
[
Menu/Hold
]
key for about 3 seconds
2) Press the
[
Set/Div
]
key for set:
-
“b-AUto”
enable
backlight when the weigh is instable;
-
“b-off”
disable
backlight
3) Press the
[
Menu/Hold
]
key for save and exit.
V. SELECT UNIT OF MEASURE
1)
Press the
[
Menu/Hold
]
key for about 3 seconds:
2)
Press the
[
M+
]
key
until of the display show
“U - xx”
;
3)
Press the
[
Set/Div
]
key for set:
-
kg (
“U - kg”
);
-
Lb (
“U - Lb”
).
4)
Press the
[
Menu/Hold
]
key for save and exit.
NOTE: It will possible see the selection unit of measure on the right of the LCD display.
VI. AUTO SWITCH-OFF FUNCTION
1)
Press the
[
Menu/Hold
]
key for about 3 seconds:
2)
Press the
[
M+
]
key
until of the display show
“Off xx”
3)
Press the
[
Set/Div
]
key for set:
-
“Off- 0”
for disable the function
Содержание OCSF-150
Страница 7: ...7 DO NOT push the load or pull the load or the loaded scale NO NO...
Страница 14: ...14 Die Last oder die Waage mit der Last d rfen NICHT geschoben oder gezogen werden NEIN...
Страница 21: ...21 NE PAS pousser ni tirer le chargement ou la balance charg e NON...
Страница 28: ...28 NON spingere n tirare il carico o la bilancia carica NO...