3.7.5.
Flow rate control settings
Parameter Possible
Value
Description
Flow rate control parameters
Min. weight variation
0 – 65535
Enable if time interval >0. If during ‘Time interval’, the net measurement
has been change of less than the ‘minimal weight variation’, then the cycle
is stopped and a flow rate error is detected.
Time Interval
0 – 65535 ms
3.7.6.
Dosing Tolerance control
Parameter Possible
Value
Description
Tolerance control parameters
Tolerance – and +
0 – 65565
Define the acceptable range for dosing results. An output assigned to the
‘out of tolerances’ is set active if the dosing result is not within the range.