5 Press the
Start
button. Y ou will see a message appear
Connecting to the USB Flash Drive
and receiv e an audible beep when the transf er of data is completed.
Note
: This method exports all data stored in the memory of the SBS-2510 into a single file that
can be accessed via Microsoft Excel.
6.1.2 Bluetooth export of results to SBS-6500 battery resistance tester
1 Insert the supplied Bluetooth dongle into the USB port of the SBS-2510.
2 Press the
Results
button on the main screen of the unit.
3 Press the
Report
button on the Results screen of the unit.
4 Under report options, Select
SBS-6500
using the Nav igation Joy stick s o print option is
selected to SBS-6500 and press the
OK
button.
5 Press the
Start
button. Y ou will see a message appear
Connecting to Bluetooth Dongle
and receiv e an audible beep when the connection is established.
6 Y ou will see a message
Initializing Bluetooth Dongle
appear and then the screen will
adv ance to
Waiting for SBS-6500
7 Press the
SG Upload
button of the SBS-6500 unit and f ollow the procedure f or importing
specif ic grav ity readings and sav ing them in the correct battery string location. (Guidance on
the operation of the SBS-6500 unit is in its own operating manual.) Below is a link f or the f ull
SBS-6500 Battery Analy zer Manual)
https://www.sbsbattery .com/media/PDFs/SBS-6500-Battery -Analy zer-Manual.pdf
Note
: All readings stored in the memory are transferred at the same time.
7 Maintenance
In this chapter y ou f ind descriptions of the maintenance tasks y ou should perf orm on y our SBS-
2510. Any other maintenance tasks need to be perf ormed by a serv ice technician that has been
qualif ied by Storage Battery Sy stems, LLC .
If y ou experience problems with y our SBS-2510, contact y our authorized Storage Battery Sy stems,
LLC Sales or serv ice representativ e.
SBS recommends that a prev entiv e maintenance and calibration certif ication is done at least once
a y ear through y our authorized SBS sales or serv ice representativ e.
https://www.sbsbattery .com/sbs -2510-digital-battery -hydrometer.html
7.1 Maintenance schedule
If the standard operating procedures of y our company require other maintenance interv als, use the
interv als listed in the standard operating procedures.
Daily
Clean the measuring cell at the end
of the work day .
Perf orm a test with deionized water.
[Check the measurement
accuracy
Frequency
Task
See
Содержание SBS-2510
Страница 1: ...SBS 2510 Digital Density Meter User Manual 1 800 554 2243 www sbsbattery com Test sbsbattery com...
Страница 2: ......
Страница 26: ...24 SBS 2510 Digital Density Meter Meter...
Страница 27: ...SBS 2510 Digital Density Meter Meter 25...
Страница 28: ...SBS 2510 Users Manual Rev 04 20 IOM2510 April 2020...