12
SBS-2510 Digital Density Meter
See also
Example: Density determination using the sample pump
5.4.2
Prepare the measuring cell for the next sample
At the end of this phase, the residue in the measuring cell must be soluble in the next sample. The
solubility of the sample def ines if y ou can drain the measuring cell or if y ou need to rinse it.
Drain
If the next sample dissolv es the sample y ou just measured, y ou can drain the cell.
Rinse
If the next sample does not dissolv e the sample y ou just measured, y ou hav e to rinse the measuring
cell with a cleaning solution. The cleaning solution must dissolv e the sample that y ou just measured
and the next sample. When y ou rinse the measuring cell, the cleaning solution remov es the residue
f rom the original sample.
See also
5.5 Exam ple: Density determ ination using the sample pump
The f ollowing chapters show y ou how to conf igure and perf orm a density determination of
deionized water at room temperature 77.0°F (25
C).
Summary of the configuration
•
Guided density determination
•
Filling and draining with the sample pump
•
Ev aluation if result lies within specif ied limits
5.5.1
Configure the method
The density unit is set to [g/cm
3
].
Confirmation of results
in
Settings
is activ ated.
The home screen is open.
1 Mov e the nav igation joy stick to the lef t.
The
Methods
window opens.
2 Press the
New
key .
The
Method settings
window opens.
3 Nav igate to
Method ID
and enter a name with up to 25 characters.
4 Press the
OK
key .
The method is listed in the
Methods
window.
5 Change the parameter settings to the v alues shown in the f ollowing table.
Parameter
Setting
Explanation
Workflow
Guided
The SBS-2510 prompts users to f ill and drain
the measuring cell.
Sample ID entry
Variable
The SBS-2510 prompts users to enter the
sample ID bef ore the density is measured.
Fill cycle
3
The SBS-2510 prompts users to f ill and drain the
measuring cell twice and then f ill the measuring
cell f or the measurement.