SBS-2510  Digital Density  Meter

 

 

5   Operation

 

5.1   Start up the SBS-2510

 

−  Press key  1 (1).

 

 

A short beep is play ed.

 

 

A welcome message is display ed.

 

 

The home  screen opens. The  SBS-2510  is ready  to use.

 

 

 

 
 

 

5.2   Shut dow n the SBS-2510

 

−  Press and hold any  of  the key s or the nav igation joy stick f or more than 3 s.

 

 

A beep is play ed and the screen goes dark.

 

 

The control circuits f or the key s and the nav igation joy stick  are energized.  The  rest of  the 

SBS-2510  is no longer energized.

 

5.3   Navigate and enter inform ation

 

5.3.1 

 Nav igate through menus and confirm settings

 

1  Press the 

Menu 

key .

 

2  To  nav igate to a desired menu, mov e the 

nav igation joy stick up or down.

 

3  To  nav igate to a submenu of  the selected menu, 

mov e the nav igation joy stick to the right.

 

4  To  go back, mov e the nav igation joy stick to the lef t.

 

 

 

 

Confirm settings

 

In ev ery  menu, where  the f unction of  key  3 is marked with a green bar, y ou can press the 
nav igation joy stick to conf irm the selection.

 

Shortcut to the method list

 

 

The  home screen is open.

 

− Mov e the nav igation joy stick to the lef t.

 

 

The 

Methods 

window  opens.

 

5.3.2 

 Enter text and numbers

 

 

The  key board window  is open.

 

1  Nav igate to the character, number, sy mbol or icon.

 

2  Press the nav igation joy stick to conf irm the selection of  the character, number, s y mbol or icon.

 

3  To  conf irm the entered inf ormation and leav e the key board window,  press the 

OK 

key .

 

Содержание SBS-2510

Страница 1: ...SBS 2510 Digital Density Meter User Manual 1 800 554 2243 www sbsbattery com Test sbsbattery com...

Страница 2: ......

Страница 3: ...and numbers 10 5 4 Typical phasesof density determinations 11 5 4 1 Fill the measuringcell andmeasure the density 11 5 4 2 Prepare the measuring cell for the next sample 12 5 5 Example Density determi...

Страница 4: ...hydrometer html If y ou hav e any additional questions contact y our authorized SBS sales representativ e or dealer Conventions and symbols Note For usef ul inf ormation about the product Ref ers to a...

Страница 5: ...e complies with Industry Canada license exempt RSS standard s and part 15 of the FCC Rules Operation is subject to the f ollowing two conditions 1 this dev ice may not cause interf erence and 2 this d...

Страница 6: ...t is not handled correctly 1 Do not store or use the battery at temperatures below 20 C or abov e 60 C 2 Do not subject the battery to mechanical stresses like pressure bending or impacts Damage to th...

Страница 7: ...if the sample pump is f illed or empty 6 Filling tube To f ill and drain the measuring cell 7 Filling opening with locking screw To f ill and drain the measuring cell using a sy ringe The locking scr...

Страница 8: ...or result and the temperature of the measuring cell 5 Sampling ribbon Display s the current status of the sample pump or sy ringe 6 Key f unctions ribbon Display s the f unction of the key s 3 2 2 Ic...

Страница 9: ...sy ringe f illing tube Washer Battery Li Ion 2400mAh 2510 TUBE 190 User Manual 4 2 Unpack the SBS 2510 1 Remov e the SBS 2510 f rom the protectiv e packaging 2 Store the packing material f or later t...

Страница 10: ...is cov ered or in a container it is not suf f iciently cooled and ov erheats 1 Do not cov er the AC adapter 2 Do not put the AC adapter in a container Assemble the AC adapter 1 Slide the desired set o...

Страница 11: ...ss the Menu key 2 Nav igate to Settings Measurement Sampling The Sampling window opens 3 Nav igate to Sample pump 4 Press the OK key 5 Nav igate back to the home screen The icon is display ed in the s...

Страница 12: ...te to a desired menu mov e the nav igation joy stick up or down 3 To nav igate to a submenu of the selected menu mov e the nav igation joy stick to the right 4 To go back mov e the nav igation joy sti...

Страница 13: ...easuring cell alway s contains residue f rom the prev ious sample or the cleaning solution it is important that y ou remov e the residue using the new sample This is only possible if the residue is so...

Страница 14: ...ters show y ou how to conf igure and perf orm a density determination of deionized water at room temperature 77 0 F 25 C Summaryof the configuration Guided density determination Filling and draining w...

Страница 15: ...to drain the measuring cell Printout None The result is not printed 5 5 2 Perform the density determination Start the method The SBS 2510 is set up to work with the sample pump The piston is in the lo...

Страница 16: ...BS 6500 6 1 Saving and exporting data Measured results are stored in the memory of the SBS 2510 Digital Density Meter These results will remain stored in memory ev en af ter transf er has been complet...

Страница 17: ...ng them in the correct battery string location Guidance on the operation of the SBS 6500 unit is in its own operating manual Below is a link f or the f ull SBS 6500 Battery Analy zer Manual https www...

Страница 18: ...ple there is a risk that the measuring cell can be damaged by the sample To prev ent such damages SBS recommends that y ou clean the measuring cell at the end of each sample series To clean the measur...

Страница 19: ...check the measurement accuracy of the SBS 2510 y ou need to conf igure and perf orm a test 7 3 1 Example Test with water The f ollowing chapters show y ou how to conf igure and perf orm a guided test...

Страница 20: ...ps when the piston is in the uppermost position 3 Hold the f illing tube ov er the waste beaker 4 Press the OK key The sample pump starts and stops when the piston is in the lowermost position 5 Clean...

Страница 21: ...ment data and repeat the test 7 If the test continues to f ail perf orm an adjustment See also Clean the measuring cell Page 16 7 4 Replace the battery The SBS 2510 works with a rechargeable 2400 mAh...

Страница 22: ...peratures of 4 122 F 20 50 C But the lif espan of the battery is reduced when it is stored at high temperatures 1 Clean the measuring cell 2 Shut down the SBS 2510 3 Clean the housing 4 Store the SBS...

Страница 23: ...Measuring cell Borosilicate glass Parts in contact when working with sample pump PTFE poly tetraf luorethy lene PVDF poly v iny lidene dif luoride EPDM ethy lene propy lene diene monomer M class rubb...

Страница 24: ...cm3 Repeatability 0 0005 g cm3 Resolution 0 0001 g cm3 Temperature Range f or sample 32 122 F 0 50 C Resolution 0 1 F 0 1 C Accuracy 0 3 F 0 2 C 1 For v iscosity range of 0 200 mPa s Measurement scal...

Страница 25: ...SBS 2510 Digital Density Meter Meter 23 Characteristic Value Brix d w w Range 0 0 85 0 w w Accuracy 0 3 w w Plato Range 0 0 85 0 P Accuracy 0 3 P...

Страница 26: ...24 SBS 2510 Digital Density Meter Meter...

Страница 27: ...SBS 2510 Digital Density Meter Meter 25...

Страница 28: ...SBS 2510 Users Manual Rev 04 20 IOM2510 April 2020...

Отзывы: