Instruction Manual 

HMR 

 

HMR-BA-e-1212 

 

6

 

 

 

 
 

                                                                                                       
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

6.5    Test Set 
 

At  the  main  display  interface, 

MENU 

key  has  to  be 

pressed to enter the main menu. 

 

 

ETR 

has  to  be  pressed  to  enter  Test  Set  Menu.  The 

symbol 

↓ at the left side of underside menu indicates that 

the  menu  has  not  ended. 

key  has  to  be  pressed  to 

glance downward continuously. 
The symbol 

↑ 

at the left side of the upside menu indicates 

that  the  menu  has  not  ended. 

has  to  be  pressed  to 

glance upward continuously. 

 

The 

or 

key  has  to  be  pressed  to  move  the 

cursor  to  the  desired  line  to  set  and 

ETR 

has  to  be 

pressed to confirm. 

Note: 1. When 

Hard/

бb

 

is switched to 

бb, hardness 

scale  cannot  be  selected.  The  cursor  will  skip  over 

Hardness Scale

while it is moved. 

2. Only D type of impact device has got the function of  
бb  measurement.  So  the  cursor  cannot  be  moved  to 

Hard/

бb

while another impact device is used. 

 

6.5.1 Impact Direction Setting 
 

 

The 

key  or  the 

key  has  to  be  pressed  to 

move the cursor to the desired impact direction. 

ETR  

has to be pressed to confirm. 

ESC  

has to be pressed to cancel. 

 

6.5.2 Average Times Setting 
 

 

Average times can be modified within the range of 1 to 32. 
The 

~     

key  has  to  be  pressed  to  input  the 

number  value.  The  cursor  will  shift  right  around 
automatically when inputting the number. 

ETR  

has to be pressed to confirm. 

ESC  

has to be pressed to cancel. 

 

6.5.3 Material Setting 
 

If 

Hard/

бb

is  preset

 

  to  hardness, following material  will 

be displayed: Steel and Cast Steel, Cold Work Tool Steel, 
Stainless  Steel,  Grey  Cast  Iron,  Nodular  Cast  Iron,    Cast 
Aluminium  Alloys,  Copper-Zinc  Alloys,  Copper-Aluminium 
Alloys, Wrought Copper and Wrought Steel. 

 

The 

or 

key  has  to  be  pressed  to  move  the 

cursor to the material desired to preset. 

ETR  

has to be pressed to confirm. 

ESC  

has to be pressed to cancel. 

 

Note:  1.

 

Presetting  the  hardness  scale,  HL  is 

automatically  recovered  after  material  presetting  has 
been changed. 
2. Please select material first and then select hardness 
scale. 
 

If 

Hard/

бb

is  preset

 

to 

бb, 

following  material  will  be 

displayed:  Mild  Steel,  High-Carbon  Steel,  Cr  Steel,  Cr-V 

M

ai

n

 d

isp

la

y

 in

te

rf

ac

Test Set 

Print Set 

Memory Manager 
System Set 
Software Info 

Impact Direc. 
Average 
Material 
Hardness Scale 
Tolerance Limit 
Hard/бb:Hard

 

(Print Set is NOT 
 available

 

 

View From No.1 
View From End 
View From No. 
Transfer 
Delete by No. 
Delete All 

Auto Save: 

Auto Print:

 

(N/A) 
Auto Delete: 
Auto Trans.: 
Key Sound: 
Warn. Sound: 
LCD Brightness 
Time Date Set

 

Содержание HMR

Страница 1: ...6 7 Memory Management 6 7 1 View from No 1 Group View from Ending Group 6 7 2 View from Selected No Group 6 7 3 Data Transfer 6 7 4 Delete by Group No 6 7 5 Delete all Data 6 7 6 Deletion Confirmation 6 8 Browsing Memory Data Groups 6 9 System Reset 6 9 1 LCD Brightness Setting 6 9 2 Date Time Setting 6 10 Software Information 6 11 System Calibration 6 12 EL Background Light 6 13 Auto Power Off 6 ...

Страница 2: ...able 1 1 No Item Qua ntity Remarks Stan dard Config uration 1 Main body 1 2 D type impact device 1 With cable 3 Standard test block 1 4 Cleaning brush I 1 5 Small support ring 1 6 Alkaline battery 4 AA size 7 Manual 1 8 Instrument case 1 9 Optio nal Config uration 11 Cleaning brush II 1 For use with G type impact device 12 Other type of impact devices and support rings Refer to Table 3 and Table 4...

Страница 3: ...RC HRA HV HS Display dot matrix LCD 128 x 64 dots Data memory max 500 groups Power supply 3V 2AA size Alkaline batteries Continuous working period about 50 hours with backlight off Communication interface USB1 1 4 Preparation Testing 4 1 Preparation and Inspection before Testing 4 1 1 Preparation of Sample Surface Preparation of the sample surface should conform to the relative requirements in APP...

Страница 4: ...b hardness tester and a corresponding hardness tester have to be used to test at the same sample respectively For each hardness value each homogeneous measurement 5 points of Leeb hardness value are needed They are in the surroundings of more than three indentations which need conversion of hardness The Leeb hardness arithmetic average value and the corresponding hardness average value as correlat...

Страница 5: ...he mean value will be displayed 6 3 3 Key Operation at the Main Display The SAVE key has to be pressed to store the present group of measured values into memory This operation is only valid after displaying the mean value The DEL key has to be pressed to delete the latest single measured value After pressing this key on the screen will be shown following The key or the key has to be pressed to mov...

Страница 6: ...ncel 6 5 2 Average Times Setting Average times can be modified within the range of 1 to 32 The 0 9 key has to be pressed to input the number value The cursor will shift right around automatically when inputting the number ETR has to be pressed to confirm ESC has to be pressed to cancel 6 5 3 Material Setting If Hard бb is preset to hardness following material will be displayed Steel and Cast Steel...

Страница 7: ...ceeds the measurement range the user will be reminded by the Tester to reset 2 If the bottom limit is larger than the upper limit they will be exchanged automatically 6 5 6 Hardness бb Setting ETR has to be pressed to switch between Hard and бb Note Only D and DC type of impact device have got the function of бb measurement So Hard is the only selection if the impact device is not D or DC type 6 6...

Страница 8: ...ssed to see details of that group The or key has to be pressed to browse details including average value test set and each single value ESC has to be pressed to return to previous display 6 9 System Set At the main display interface MENU key has to be pressed to enter the main menu or key has to be pressed to move the cursor to System Set Menu ETR has to be pressed to enter System Set Menu Auto Sa...

Страница 9: ... if it hasn t been used for a longer time or after having reset the system The key and the ETR key have to be pressed to power on the system Then the software calibration screen will display following The impact direction has to be set at 5 points have to be measured on the standard hardness block The average value will be displayed after measuring 5 times or key has to be pressed to input the nom...

Страница 10: ...e unscrewed first Then the impact body has to be taken out The nylon brush has to be spiralled in counter clock direction into the bottom of the guide tube It has to be taken out and repeated for 5 times At last the impact body and the support ring have to be installed again The impact body has to be released after use Any lubricant is absolutely prohibited inside the impact device 8 2 Normal Main...

Страница 11: ... 7 68 2 22 6 70 2 HV 80 898 80 935 100 941 82 1009 Stainless steel HRB 46 5 101 7 HB 85 655 HV 85 802 Grey cast iron HRC HB 93 334 92 326 HV Nodular cast iron HRC HB 131 387 127 364 HV Cast aluminium alloys HB 19 164 23 210 32 168 HRB 23 8 84 6 22 7 85 0 23 8 85 5 BRASS copper zinc alloys HB 40 173 HRB 13 5 95 3 BRONZE copper alum tin alloys HB 60 290 Wrought copper alloys HB 45 315 Table 2 No Mat...

Страница 12: ...1 6μm Min weight of sample Measure directly Need support firmly Need coupling tightly 5kg 2 5kg 0 05 2kg 5kg 2 5kg 0 05 2kg 1 5kg 0 5 1 5kg 0 02 0 5kg 15kg 5 15kg 0 5 5kg 5kg 2 5kg 0 05 2kg Min Thickness of sample Coupling tightly Min layer thickness for surface hardening 5mm 0 8mm 5mm 0 8mm 1mm 0 2mm 10mm 1 2mm 5mm 0 8mm Size of tip indentation Hardness 300HV Indentation diameter Depth of indenta...

Отзывы: