Instruction Manual
HMR
HMR-BA-e-1212
5
screen will exit and the main display interface will be
entered.
6.2 Power Off
By pressing the
key the system can be powered off in
any state.
6.3 Testing
Below, the main display is shown:
6.3.1 Full description of the Main Display
Battery information:
displays the rest capacity of the
battery.
Impact direction:
displays the present impact direction.
Average value indicator:
it appears to show the mean
value when the preset impact times are reached.
Hardness scale:
Hardness method of the present
measured value.
Measured value:
displays the present single time
measured value (without mean value indicator), or displays
the present mean value (with average value indicator
prompting).
↑
means that the value is over conversion or
measurement range.
↓ means that it is lower than
conversion or measurement range.
Material:
The present preset material.
Impact times count:
Times that have been impacted.
6.3.2 Testing Operation at the Main Display
Testing operation can be carried out under this interface.
After each impact operation, the current measured value
can be displayed. Impact times count plus one, the buzzer
will alert a long howl if the measured value is not within the
tolerance limit. As soon as the preset impact times are
reached, the buzzer will alert a long howl. After two
seconds, the buzzer will alert a short howl and the mean
value will be displayed.
6.3.3 Key Operation at the Main Display
- The
【
SAVE
】
key has to be pressed to store the present
group of measured values into memory. This operation is
only valid after displaying the mean value.
- The
【
DEL
】
key has to be pressed to delete the latest
single measured value. After pressing this key, on the
screen will be shown following:
The
key or the
key has to be pressed to
move the cursor to
【
YES
】
or
【
NO
】
.
【
ETR
】
has to be pressed to confirm.
【
ESC
】
has to be pressed to cancel deletion.
- The
【
】
or
【
】
key has to be pressed to display the
single measured value.
- By pressing
【
AVG , testing can be ended although the
preset impact times have not been reached and to display
the average value.
- The background light of LCD has to be switched on and
off by pressing
【
】
.
-
【
MENU
】
has to be pressed to enter the system Preset
Menu.
- By pressing
, the impact direction can be preset.
-
【
TIME
】
key has to be pressed to change the impact
times in one group. The impact times count item will be
highlighted when the
【
TIME
】
key has been pressed. The
impact times count value will be added one with each
pressing. The value will roll back to 1 after 32 has been
reached.
- The
【
HD
】
key has to be pressed to change the hardness
scale.
- By pressing
【
MTL
】
material setting can be changed.
Presetting the hardness scale, HL is automatically
recovered after material presetting has been changed.
6.4 Menu Structure
Both, presetting system parameters and additional
functions, can be realised by menu operation. At the main
display interface,
【
MENU
】
key has to be pressed to enter
the main menu, as shown in following picture:
Ma-
terial
Average value
indicator
Hardness
scale
Meas
ured
value
Impact times
count
Impact
direction
Battery
Inform.