Instruction Manual

 

FH

 

FH Adjustment Procedure 

1

  

Start the 

Press ON/OFF

 

instrument

 

2

  

Enter the 

Directly after pressing 

adjustment 

ON/OFF, press PEAK 

procedure 

and PRINT together for 
more times and 

very 

short intervals 

until the 

red light is on. 

3  Instrument 

Press SET 

Type 

Directly after pressing 
PEAK + PRINT 

a

  

(Back to the 

(If you are in the 

normal 

operation mode again, 

mode ???) 

the instrument has to be 
shut off. Start again with 
1. Eventually pressing 
the buttons faster) 

4

  

Select the 

Press 

▼▲

 

instrument

 

type

 

5

  

Save 

Press SET

 

instrument

 

type

 

6

  

Enter the 

Press MEMORY

 

adjustment

 

sequence

 

7

  

Choose the 

Press UNIT and

 

adjustment 

with

▼▲ specify the

 

weight 

calibration weight in 

Newton (X kg * 9,81) 

8

  

Save the 

Press SET and UNIT 

adjustment 

simultaneously

 

weight

 

9

  

Attach the 

Press ZERO 

weight to the

 

instrument

 

(by hanging)

 

finished 

The green 
light is on 

The lower 
red light is 
on 

XXX N is 
shown 

The 
correspondi 
ng value to 
the instr. 
appears in 
the display 

The upper 
red light is 
on 

The weight 
is written in 
the display 

The 
instrument 
changes to 
operation 
mode. 
Adjustment 
procedure is 

FH_SI-BA-e-1317 

Содержание FH-S

Страница 1: ...cl rechargeable battery Carrying Case Charger Standard Attachments as shown below 5 pcs M3 x 8 screws Power plug Reset button Display Pilot lamp RS 232 port Force Sensor Sensor Inside Important annotation By pressing the RESET key on the right side of housing see illustration on the left individual settings and memorised values can be re set or erased in example for a new start of the instrument a...

Страница 2: ...ent off after a here defined time 9 AVERAGE or Saving Mode period in sec P OFF To change press or Only active in Battery Mode Operation keys ON OFF 1 x Press Peak Freeze Time PE 2E ON OFF key A PE HoldT Time period in sec in For ON press 1 sec which a peak value is being shown in the display in sec To change press or UNIT 1 x Press Data Transfer rS232 to PC Press shortly Select unit P OFF HoldT St...

Страница 3: ...ound To set these limit values please see the SET Menu in section Operating keys Measurement Track Modus Display 1 shows the continuous force in a defined direction 6 To zero the display press Peak Hold Function Peak Mode Please press Auto Peak Hold Function Auto Peak Mode Please press Min limit peak function to activate storage of measurements This function allows to eliminate unwanted Pre Peak v...

Страница 4: ...material to be measured incorrect measuring results can be displayed due to the stability compensation in the instrument Example Slow draining off of liquid from a container suspended from the instrument Do not attach a continuous load This can damage the measuring unit as well as the parts relevant to safety Prevent jolts torsion and oscillation e g by appending slopingly of all kinds Be sure to ...

Страница 5: ...in KERN s accredited DKD calibration laboratory return to national normal Fundamental safety information Do not use the hanging instrument to transport loads Prevent jolts torsion and oscillation e g by appending slopingly of all kinds Never use the hanging instrument over the maximum permitted weight Danger of breaking Always make sure that there are no living beings or materials below the load t...

Страница 6: ...he instrument has to be shut off Start again with 1 Eventually pressing the buttons faster 4 Select the Press instrument type 5 Save Press SET instrument type 6 Enter the Press MEMORY adjustment sequence 7 Choose the Press UNIT and adjustment with specify the weight calibration weight in Newton X kg 9 81 8 Save the Press SET and UNIT adjustment simultaneously weight 9 Attach the Press ZERO weight ...

Отзывы: