Instruction Manual
FH
FH Adjustment Procedure
1
Start the
Press ON/OFF
instrument
2
Enter the
Directly after pressing
adjustment
ON/OFF, press PEAK
procedure
and PRINT together for
more times and
very
short intervals
until the
red light is on.
3 Instrument
Press SET
Type
Directly after pressing
PEAK + PRINT
a
(Back to the
(If you are in the
normal
operation mode again,
mode ???)
the instrument has to be
shut off. Start again with
1. Eventually pressing
the buttons faster)
4
Select the
Press
▼▲
instrument
type
5
Save
Press SET
instrument
type
6
Enter the
Press MEMORY
adjustment
sequence
7
Choose the
Press UNIT and
adjustment
with
▼▲ specify the
weight
calibration weight in
Newton (X kg * 9,81)
8
Save the
Press SET and UNIT
adjustment
simultaneously
weight
9
Attach the
Press ZERO
weight to the
instrument
(by hanging)
finished
The green
light is on
The lower
red light is
on
XXX N is
shown
The
correspondi
ng value to
the instr.
appears in
the display
The upper
red light is
on
The weight
is written in
the display
The
instrument
changes to
operation
mode.
Adjustment
procedure is
FH_SI-BA-e-1317
6