3.6 Optional probes and cables for KT 320 data loggers
All the probes for the
KT 320
data loggers have the
smart plug
technology.
An automatic recognition and the adjustment parameters storage make them 100% interchangeable.
Part No.
Description
Measuring ranges
Accuracies*
Image & dimensions
External or ambient thermo-hygrometric probes
KITHA
Interchangeable
hygrometry and
ambient
temperature
probe
0 to 100%RH
-4 to 158°F
(-20 to 70°C)
±2%RH from 5 to 95%RH @ 59°F to 77°F (15 to 25°C)
Temp: ±0.8°F from 32 to 122°F
(±0.4°C from 0 to 50°C)
±1.5°F below 32°F or above 122°F
(±0.8°C below 0°C or above 50°C)
3.7'' (94.5 mm)
KITHP-130
Remote
interchangeable
hygrometry and
temperature
probe
0 to 100%RH
-4 to 158°F
(-20 to 70°C)
±2%RH from 5 to 95%RH @ 59°F to 77°F (15 to 25°C)
Temp: ±0.8°F from 32 to 122°F
(±0.4°C from 0 to 50°C)
±1.5°F below 32°F or above 122°F
(±0.8°C below 0°C or above 50°C)
5.12'' (130 mm)
KITHI-150
Remote
interchangeable
hygrometry and
temperature
probe
0 to 100% RH
-40 to 356°F
(-40 to 180°C)
±1.5%RH from 5 to 95%RH @ 59 to 77°F (15 to 25°C)
±0.6°F from -40 to 32°F
±(0.2% of r0.6°F) from 32 to 356°F
(±0.3% of reading ±0.25°C)
5.9'' (150 mm)
General use or insertion Pt100 temperature probes
KIRGA-50
Immersion probe
(IP 65)
-40 to 248°F
(-40 to 120°C)
±0.6°F from -40 to 32°F
±(0.3% of r0.5°F) from 32 to 248°F
(±(0.4% of the r0.3°C) from -40 to 120°C)
Ø0.24'' x 1.97'' (Ø6 x 50 mm)
KIRGA-150 Immersion probe
(IP 65)
-40 to 248°F
(-40 to 120°C)
±0.6°F from -40 to 32°F
±(0.3% of r0.5°F) from 32 to 248°F
(±0.4% of the r0.3°C from -40 to 120°C)
Ø0.24 x 5.9'' (Ø6 x 150 mm)
KIRAM-150
Ambient probe
-40 to 248°F
(-40 to 120°C)
±0.6°F from -40 to 32°F
±(0.3% of r0.5°F) from 32 to 248°F
(±0.4% of the r0.3°C from -40 to 120°C)
Ø0.24 x 5.9'' (Ø6 x 150 mm)
KIRPA-150 Penetration probe
(IP 65)
-58 to 482°F
(-50 to 250°C)
±0.6°F from -58 to 32°F
±(0.3% of r0.5°F) from 32 to 482°F
(±0.4% of the r0.3°C from -50 to 250°C)
Ø0.24 x 5.9'' (Ø6 x 150 mm)
KIPI3-150-E
Penetration probe
with handle (IP
68)
-58 to 482°F
(-50 to 250°C)
±0.6°F from -58 to 32°F
±(0.3% of r0.5°F) from 32 to 482°F
(±0.4% of the r0.3°C from -50 to 250°C)
Ø0.12'' x 5.9'' (Ø3 x 150 mm)
KICA-320 Smart adapter for
Pt100 probe
-328 to 1112°F
according to the probe
(-200 to 600°C)
N/A
Ø0.12'' x 5.9'' (Ø3 x 150 mm)
Input current, voltage and impulsion cables
KICT
Voltage input
cable
0 to 5V
or 0 to 10V
±0.2% of the measurement ±1mV
KICC
Current input
cable
0 to 20 mA
or 4 to 20 mA
±0.2% of the measurement ±1µA
KICI
Pulse input cable
Maximal voltage: 5V
Type of input: TTL frequency counting
Maximal frequency: 10 KHz
* All accuracies specified in this document were conducted under laboratory conditions and can be guaranteed for measurements carried out in the same conditions, or carried out with compensation.
TECHNICAL FEATURES
11