. Calibration mode 

By performing the 3-point calibration, this instrument can accurately measure the pH values. It is recommend

 

more accurate measurement, calibration of about once a day is recommended. 

 

Before calibration, prepare the following items: 

  pH 4 standard solution (phthalate solution): pH 4.01 (at 25°C) 

  pH 7 standard solution (neutral phosphate solution): pH 6.86 (at 25°C) 

  pH 10 standard solution (carbonate solution): pH 10.01 (at 25°C) 

  Distilled water, pure water or tap water 

  Beakers (each beaker to contain one type of standard solution and water (distilled, pure or tap water) 

     

     

  * Prepare the beakers for water as many as the standard solutions. 

 
These standard solutions are separately purchased.    Contact us or the shop where you purchased the 
SK-670PHII. 
 
(1) How to calibrate the SK-670PH pH meter 

  Press the PWR key to turn on.   

  Remove the glass electrode protection bottle and wash the glass electrode thoroughly with water 

     

(distilled, pure or tap water) and wipe clean the water on the glass electrode. After washing, use clean

 

filter paper, cotton swabs or tissue paper to absorb the water off the glass electrode. 
NB:  When washing the glass electrode, be careful not rub it. Doing so could generate static electricity 

resulting in unstable pH readings. 

  Immerse the glass electrode into a standard solution.    Approx. pH value will be displayed. Then, 

   

slowly stir the solution with the electrode, and then hold the unit still.   

 

* For more accurate pH measurement, perform the calibration on pH 7, pH 4 and pH 10 in due order. 

  Press the CAL key to enter the calibration mode.    The unit automatically recognizes the calibration point 

   

and the display the pH value of the calibration point.    The current liquid temperature value is displayed 

     

at same time.   

 

                                                                        When the liquid temperature of pH 7 standard solution is 25°C   

 

 

 

 

The table below shows the pH value range to be set for calibration. 

Calibration point 

Default value 

Setting range 

pH 4 

4.00 

3.50 to 4.50 

pH 7 

7.00 

6.50 to 7.50 

pH 10 

10.00 

9.50 to 10.50 

 

Содержание 6433-00

Страница 1: ...Instruction Manual No 6433 00 Pen Type pH meter Model SK 670PH P08...

Страница 2: ......

Страница 3: ...t and may malfunction Do not leave the unit in such a place If this unit is used in an environment where electrical noise is generated the display may become unstable or the measurement error may incr...

Страница 4: ...115mm L that can be directly inserted into a flask or like at measurement Large LCD for easy reading The pH value and temperature are indicated on a large liquid crystal display LCD United the display...

Страница 5: ...lectrode protection bottle Contains the phthalate pH standard solution pH4 to protect and moisture retention of glass electrode 6 Liquid junction 7 Glass electrode 8 Temperature sensor 9 Battery cover...

Страница 6: ...proof purpose the lid of the battery compartment is fastened very tight Be careful not to hurt yourself by using your fingernails to unfasten Also be careful not to have any dust on the connecting are...

Страница 7: ...ly wash the contaminated skin with running water If the solution gets into the eyes immediately rinse with water and consult a doctor If the standard solution in the bottle runs down replenish the bot...

Страница 8: ...ectrode thoroughly with water distilled pure or tap water and wipe clean the water on the glass electrode After washing use clean filter paper cotton swabs or tissue paper to absorb the water off the...

Страница 9: ...value to calibration value refer to the Relation between pH and temperature After 10 sec pH value is fixed and the value has been stored in the unit memory The calibration is ended and automatically...

Страница 10: ...pH value according to the temperature of the standard solution as follows Standard solution Name of solution Temp pH value Temp pH value pH 4 Phthalate solution 0 C 4 00 30 C 4 02 5 C 4 00 35 C 4 02...

Страница 11: ...ced with a new one if the electrode is broken or the performance is degraded The glass electrode is not repairable Hold function Use the HOLD function when the temperature of the sample being measured...

Страница 12: ...e or protection bottle this does not pose a problem when using the instrument The deposit is easily washed off with tap water If it does not come off with tap water soak the electrode in warm water to...

Страница 13: ...eplace the sensor with a new one No calibration was performed Perform a calibration The sample amount is not enough Immerge the sensor into the sample 3cm deep from the distal end of the sensor The me...

Страница 14: ...t Coin lithium Battery CR2031 x 2 pcs Battery life About 400 hours continuous Materials Body ABS resin Electrode Glass Dimensions Body including sensing section W 36 H 224 D 20 mm Sensing section 12 d...

Страница 15: ......

Страница 16: ...3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan URL http www sksato co jp english...

Отзывы: