Installing and replacing batteries 

Install or replace batteries when firstly use of the unit or [BAT] mark is lit.   
(1)  Remove the battery cover by turning it clockwise with a coin or like. 

(2)  When replacing the batteries, remove the old batteries. 

(3)  Install new batteries with attention of the polarity marks (+ and -)   

(4)  Place batteries overlapping two batteries with the ‘+’ mark facing up.   

(5)  Reinstall the cover and tighten it by

 

turning it counterclockwise 

with a coin or like   

 

 

Cautions

 

-  The batteries are factory preset and since those batteries are for monitoring purposes, they may not last as 

long as general batteries. 

-  When the low battery mark blinks, replace new batteries. If you continue to use the unit with the mark 

blinking, malfunctions can occur. 

-  Do not dispose of used batteries in fire. 

-  Keep the batteries out of reach of children. If swallowed accidentally, consult a medical attention 

immediately. 

-  For environmental protection, dispose of used battery in compliance with local rules and regulations. 

-  For waterproof purpose, the lid of the battery compartment is fastened very tight. Be careful not to hurt 

yourself by using your fingernails to unfasten. 

-  Also be careful not to have any dust on the connecting area. If it is dirty, clean well before fastening. 

-  Do not let the unit get wet when the battery is removed. If water gets inside the unit, the instrument might 

get damaged. 

 

Before Measuring 

There is phthalate solution inside the cover avoiding the glass electrodes becomes dry.    Be careful not to 
spill when removing.    After the measurement, put the cover on the glass electrodes 

NB:

 When the unit is used for the first time or after it has been stored for several days, the pH measurement 

response may be slower. In such cases, immerse the sensor probe electrode in either the pH standard 
solution or water and let it stand for one hour. For more accurate measurement, immerse it 12 hours at 
least in the pH 4 standard solution before use.    (Turn off the unit during this immersion time.) 

 

(1) Removing the glass electrode protection bottle 
 

 Hold the sensor probe upright so as not to spill the standard solution in 

the protection bottle. Turn the protection bottle counterclockwise while 
pressing the cap. 

 
 
 

"

Protection bottle

Cap

Содержание 6433-00

Страница 1: ...Instruction Manual No 6433 00 Pen Type pH meter Model SK 670PH P08...

Страница 2: ......

Страница 3: ...t and may malfunction Do not leave the unit in such a place If this unit is used in an environment where electrical noise is generated the display may become unstable or the measurement error may incr...

Страница 4: ...115mm L that can be directly inserted into a flask or like at measurement Large LCD for easy reading The pH value and temperature are indicated on a large liquid crystal display LCD United the display...

Страница 5: ...lectrode protection bottle Contains the phthalate pH standard solution pH4 to protect and moisture retention of glass electrode 6 Liquid junction 7 Glass electrode 8 Temperature sensor 9 Battery cover...

Страница 6: ...proof purpose the lid of the battery compartment is fastened very tight Be careful not to hurt yourself by using your fingernails to unfasten Also be careful not to have any dust on the connecting are...

Страница 7: ...ly wash the contaminated skin with running water If the solution gets into the eyes immediately rinse with water and consult a doctor If the standard solution in the bottle runs down replenish the bot...

Страница 8: ...ectrode thoroughly with water distilled pure or tap water and wipe clean the water on the glass electrode After washing use clean filter paper cotton swabs or tissue paper to absorb the water off the...

Страница 9: ...value to calibration value refer to the Relation between pH and temperature After 10 sec pH value is fixed and the value has been stored in the unit memory The calibration is ended and automatically...

Страница 10: ...pH value according to the temperature of the standard solution as follows Standard solution Name of solution Temp pH value Temp pH value pH 4 Phthalate solution 0 C 4 00 30 C 4 02 5 C 4 00 35 C 4 02...

Страница 11: ...ced with a new one if the electrode is broken or the performance is degraded The glass electrode is not repairable Hold function Use the HOLD function when the temperature of the sample being measured...

Страница 12: ...e or protection bottle this does not pose a problem when using the instrument The deposit is easily washed off with tap water If it does not come off with tap water soak the electrode in warm water to...

Страница 13: ...eplace the sensor with a new one No calibration was performed Perform a calibration The sample amount is not enough Immerge the sensor into the sample 3cm deep from the distal end of the sensor The me...

Страница 14: ...t Coin lithium Battery CR2031 x 2 pcs Battery life About 400 hours continuous Materials Body ABS resin Electrode Glass Dimensions Body including sensing section W 36 H 224 D 20 mm Sensing section 12 d...

Страница 15: ......

Страница 16: ...3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan URL http www sksato co jp english...

Отзывы: