SATO 6433-00 Скачать руководство пользователя страница 10

 

 
 

 

  The figure above shows the case when the pH 10 standard solution is 25.0 ºC 

 
 
 
 
 
When the sample is known as acidic or alkaline, some of the calibration can be skipped 
 
Acidity: Perform calibration on pH7 and pH 4 and skip on pH10 
Alkalinity: Perform calibration on pH 7 and 10 and skip on pH 4 

 
 
 

Relation between pH value and temperature 
 

When performing calibration, set the pH value according to the temperature of the standard solution as follows; 

Standard solution 

Name of solution 

Temp. 

pH value

Temp. 

pH value

pH 4 

Phthalate solution 

0 ºC 

4.00 

30ºC 

4.02 

5 ºC 

4.00 

35ºC 

4.02 

10ºC 

4.00 

40ºC 

4.04 

15ºC 

4.00 

45ºC 

4.05 

20ºC 

4.00 

50ºC 

4.06 

25ºC 

4.01 

pH 7 

Neutral phosphate solution 

0 ºC 

6.98 

30ºC 

6.85 

5 ºC 

6.95 

35ºC 

6.84 

10ºC 

6.92 

40ºC 

6.84 

15ºC 

6.90 

45ºC 

6.83 

20ºC 

6.88 

50ºC 

6.83 

25ºC 

6.86 

pH 10 

Carbonate solution 

0 ºC 

10.32 

30ºC 

9.97 

5 ºC 

10.24 

35ºC 

9.92 

10ºC 

10.18 

40ºC 

9.89 

15ºC 

10.12 

45ºC 

9.86 

20ºC 

10.06 

50ºC 

9.83 

25ºC 

10.01 

                                                                                                                                                       

  JIS Z 8802 (2011)     

   

Содержание 6433-00

Страница 1: ...Instruction Manual No 6433 00 Pen Type pH meter Model SK 670PH P08...

Страница 2: ......

Страница 3: ...t and may malfunction Do not leave the unit in such a place If this unit is used in an environment where electrical noise is generated the display may become unstable or the measurement error may incr...

Страница 4: ...115mm L that can be directly inserted into a flask or like at measurement Large LCD for easy reading The pH value and temperature are indicated on a large liquid crystal display LCD United the display...

Страница 5: ...lectrode protection bottle Contains the phthalate pH standard solution pH4 to protect and moisture retention of glass electrode 6 Liquid junction 7 Glass electrode 8 Temperature sensor 9 Battery cover...

Страница 6: ...proof purpose the lid of the battery compartment is fastened very tight Be careful not to hurt yourself by using your fingernails to unfasten Also be careful not to have any dust on the connecting are...

Страница 7: ...ly wash the contaminated skin with running water If the solution gets into the eyes immediately rinse with water and consult a doctor If the standard solution in the bottle runs down replenish the bot...

Страница 8: ...ectrode thoroughly with water distilled pure or tap water and wipe clean the water on the glass electrode After washing use clean filter paper cotton swabs or tissue paper to absorb the water off the...

Страница 9: ...value to calibration value refer to the Relation between pH and temperature After 10 sec pH value is fixed and the value has been stored in the unit memory The calibration is ended and automatically...

Страница 10: ...pH value according to the temperature of the standard solution as follows Standard solution Name of solution Temp pH value Temp pH value pH 4 Phthalate solution 0 C 4 00 30 C 4 02 5 C 4 00 35 C 4 02...

Страница 11: ...ced with a new one if the electrode is broken or the performance is degraded The glass electrode is not repairable Hold function Use the HOLD function when the temperature of the sample being measured...

Страница 12: ...e or protection bottle this does not pose a problem when using the instrument The deposit is easily washed off with tap water If it does not come off with tap water soak the electrode in warm water to...

Страница 13: ...eplace the sensor with a new one No calibration was performed Perform a calibration The sample amount is not enough Immerge the sensor into the sample 3cm deep from the distal end of the sensor The me...

Страница 14: ...t Coin lithium Battery CR2031 x 2 pcs Battery life About 400 hours continuous Materials Body ABS resin Electrode Glass Dimensions Body including sensing section W 36 H 224 D 20 mm Sensing section 12 d...

Страница 15: ......

Страница 16: ...3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan URL http www sksato co jp english...