 Slide the cap to take it out from the sensor probe as shown in the figure 

on the right. 

 
 
 
 
 
 
 
 

(2) Installing the glass electrode protection bottle 

 

 Install the protection bottle to the sensor probe in the reverse order of removal. 

 

 Hold the sensor probe upright so as not to spill the standard solution in the 

protection bottle. Turn the protection bottle clockwise while pressing the cap. 

 

 After installing the protection bottle, push in the sensor probe until its tip hits the 

bottom of the protection bottle. 

  Note:  Pay attention not to spill the standard solution in the bottle when installing or removing the glass 

electrode protection bottle 

              * If the standard solution comes in contact with the skin (hand), immediately wash the contaminated 

skin with running water.    If the solution gets into the eyes, immediately rinse with water and consult a 
doctor. 

 

* If the standard solution in the bottle runs down, replenish the bottle with the phthalate pH standard 

solution to protect the glass electrode from becoming dry. Adding distilled water, pure water or tap 
water before storing is recommended. 

 

 

. Auto Power-Off Function

 

When the unit is untouched for about 20 minutes, the auto power-off function is triggered. If you need a 
continuous measurement, make sure to disable the function.     

  Release the Auto Power-off function 

      Press the PWR key while pressing on the HOLD / 

  key.     

      When “

n

” character appears on the LCD, release the keys.   

      Then the Auto power-off function is released.                   

 
 

Cautions 

When the unit gets turned off, disabled auto power-off function is no longer effective.    Each time the power 
is turned on, disable the auto power-off function.

 

 

     
 
 

Содержание 6433-00

Страница 1: ...Instruction Manual No 6433 00 Pen Type pH meter Model SK 670PH P08...

Страница 2: ......

Страница 3: ...t and may malfunction Do not leave the unit in such a place If this unit is used in an environment where electrical noise is generated the display may become unstable or the measurement error may incr...

Страница 4: ...115mm L that can be directly inserted into a flask or like at measurement Large LCD for easy reading The pH value and temperature are indicated on a large liquid crystal display LCD United the display...

Страница 5: ...lectrode protection bottle Contains the phthalate pH standard solution pH4 to protect and moisture retention of glass electrode 6 Liquid junction 7 Glass electrode 8 Temperature sensor 9 Battery cover...

Страница 6: ...proof purpose the lid of the battery compartment is fastened very tight Be careful not to hurt yourself by using your fingernails to unfasten Also be careful not to have any dust on the connecting are...

Страница 7: ...ly wash the contaminated skin with running water If the solution gets into the eyes immediately rinse with water and consult a doctor If the standard solution in the bottle runs down replenish the bot...

Страница 8: ...ectrode thoroughly with water distilled pure or tap water and wipe clean the water on the glass electrode After washing use clean filter paper cotton swabs or tissue paper to absorb the water off the...

Страница 9: ...value to calibration value refer to the Relation between pH and temperature After 10 sec pH value is fixed and the value has been stored in the unit memory The calibration is ended and automatically...

Страница 10: ...pH value according to the temperature of the standard solution as follows Standard solution Name of solution Temp pH value Temp pH value pH 4 Phthalate solution 0 C 4 00 30 C 4 02 5 C 4 00 35 C 4 02...

Страница 11: ...ced with a new one if the electrode is broken or the performance is degraded The glass electrode is not repairable Hold function Use the HOLD function when the temperature of the sample being measured...

Страница 12: ...e or protection bottle this does not pose a problem when using the instrument The deposit is easily washed off with tap water If it does not come off with tap water soak the electrode in warm water to...

Страница 13: ...eplace the sensor with a new one No calibration was performed Perform a calibration The sample amount is not enough Immerge the sensor into the sample 3cm deep from the distal end of the sensor The me...

Страница 14: ...t Coin lithium Battery CR2031 x 2 pcs Battery life About 400 hours continuous Materials Body ABS resin Electrode Glass Dimensions Body including sensing section W 36 H 224 D 20 mm Sensing section 12 d...

Страница 15: ......

Страница 16: ...3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan URL http www sksato co jp english...

Отзывы: