4.5.13 Parameter Input
Open the menu via -[Calibration]-[Param]
WP A/Calibration
Measuretime
160 ms
Digital filter
bessel
Fcut
3.00 Hz
This line is shown only, if the filter is on.
Test mode
absolute
W
&
M
none
Standstill time
0.50 s
Standstill range
1.00 d
[Measuretime]
Select the measurement time; possible values: 5, 10, 20, 40, 80, 160, 320, 640, 960, 1280, 1600 ms. In
'legal-for-trade' mode select <= 1 s.
[Digital filter]
The digital filter can be switched on only with the measurement time set to <= 160 ms.
Select the filter characteristic: [off], [bessel], [aperiod.], [butterw.], [tscheby.].
[Fcut]
Enter the cut-off frequency for the low-pass filter (0.1 – 80 Hz); the setting is dependent on the
measurement time.
[Test mode]
With [Absolute], the test value is determined when the test is called. With [Relative], the deviation from the
initially stored test value is displayed; see Chapter 4.5.11.
[W
&
M]
Setting for 'legal-for-trade' mode; select [None], [OIML], [NTEP] (for USA) or [NSC] (for Australia); see
[Standstill time]
Parameters [Standstill range] and [Standstill time] are required for determining the mechanical stability of
the scale. Input in seconds; permissible range: 0.01 to 2 sec. If 0 is set, stability is not checked. The stability
time must not be less than the measurement time.
[
Standstill range
]
The mechanical stability of the scale can be detected as long as any changes in the weight value are within
this range; permissible range 0.01d to 10.00d. In ‘legal-for-trade’ mode, select <= 1 d.